37
38
Display
What it means
Battery is installed incorrectly;
Low battery power;
Replace the battery;
Test strip is inserted incorrectly or
not completely in;
Meter does not turn
on after inserting a
test strip
No display.
Test strip port malfunction
Call Sinocare customer service at
+86-731-89935581/
+86-731-89935582
or contact your local dealers.
Remove the battery, wait for 3
minutes, reinstall the battery. If still
no display, call Sinocare customer
service at +86-731-89935581/
+86-731-89935582 or contact your
local dealers.
Check that the battery is correctly
installed with the right position of
sign “+” and “-;”
Insert the test strip correctly with the
printing facing up and push it into
strip port until it stops;
What to do
Display
What it means
Not enough blood sample
Sample is applied to wrong area
of test strip
Meter doesn’t begin
test after applying a
blood sample
Used test strip
Repeat the test with a new test strip
with blood sample drawn into right
area Please refer to “Blood
glucose testing” section.
Repeat the test with a new test strip
Test strip exposed in air too long
after opening the vial or package
The end of test strip may press
too firmly to your finger, blood
sample cannot be drawn into test
strip reaction chamber;
Use the side of test strip reaction
chamber gently touch against the
fingertip to draw a drop of blood;
Finish the test immediately after
taking out a test strip from a vial
or package;
Repeat the test with a new test strip
with enough sample
What to do
Содержание Safe-Accu