Performing a control solution test
Start with the blood glucose meter off. Prepare the following items before control solution
test:
a. blood glucose meter;
b. blood glucose test strip;
c. blood glucose control solution
STEP 1: Wash your Hands
Wash your hands with mild soap and water. Be sure to dry your hands before performing
a test.
STEP 2: Insert a Safe-Accu 2 Test Strip
Insert a Safe-Accu 2 Test Strip with printing face up and the contact end into the test
strip port until it cannot go any further. The blood glucose meter will turn on.
Attention: Only when meter display flashing blood drop symbol “ ”, can you apply
control solution. If there is no blood drop symbol, restart the meter again.
STEP 3: Apply the Control Solution
1. Check the expiration date and discard dates on your control solution and Safe-Accu 2
test strip vials. Do not use control solution or test strip which expired.
2. Shake the control solution bottle well, then remove cap. Squeeze the bottle and discard
the first drop. Squeeze the bottle again to get a second drop and bring the tip of the
Safe-Accu 2 test strip to touch the drop of solution until the blood glucose meter beeps.
After 10 seconds’ count-down, meter will display control solution test result.
3. Control test results must be within range of the control solution level printed on test
strip vial label. If the result is out of range, test again. If still so, do not use test strips or
meter for blood glucose testing. Call customer service at
+86-731-8986-731-89935582 or contact local dealers for help.
4. Eject the test strip
Attention: Discard used test strip and control solution according to local regulations.
Keep control solution away from children and pets.
Questionable Results:
Probable Cause
Operation mistake
Control solution not being shaken well
Retest according to correct steps
Shake control solution well and retest with a new test strip.
Use the first drop of control solution
Retest with a new test strip
Expired or contaminated control solution
Change a new vial control solution and retest.
Expired or contaminated test strip
Change a new vial test strip and retest.
Temperature is too high or too low
Equilibrate meter, test strip and control solution in room
temperature for 30 minutes and retest.
Meter malfunction
Contact customer service or local dealers
Action