Simpson M245 Series Скачать руководство пользователя страница 2

DISPLAY
Type:

7-segment LCD

Height:

0.5” (12.7mm)

Decimal point:

4-position programmable

Overrange indication:

Most significant digit = “1”

Backlighting:

Optional negative image, red LED

backlighting

Polarity:

Auto with “-” indication, “+” implied

POWER REQUIREMENTS
AC Volt:

85-250VAC @40-440Hz

DC Volt:

9-32VDC

Power Consumption:

85-250VAC:  2.5VA min/4VA max
9-32VDC:  1.5VA min/3VA max

ACCURACY @ 25°C

±(0.02% of input + 1 count)

ENVIRONMENTAL
Operating Temperature:

0 to 55°C

Storage Temperature:

-10 to 60°C

Relative Humidity:

0 to 85% non condensing

Temperature Coefficient:

All inputs ±(0.02% of input ±0.2 digit)/°C

Warmup time:

Less than 20 minutes

NOISE REJECTION
NMRR:

60dB, 50/60Hz

CMRR:

(w/1K

unbalanced @ 60Hz): 90dB min

ANALOG TO DIGITAL CONVERSION
Technique:

Integrating

Rate:

3 samples/second-typical

MECHANICAL
Bezel:

0.95” x 2.84”

(24mm x 72mm)

MECHANICAL (cont.)
Depth:

2.36” (60mm)

Panel cutout:

0.88” x 2.68”

(22.2mm x 68mm)

Weight:

3.5oz (99.2g)

Case Material:

94-0,UL-rated, glass-filled thermoplastic

Specifications

Wiring Display

Display Scaling

Span Adjustment:

Mini-Max indicators have limited

range coarse and fine adjustments for display scaling.
There are no optional connections required for these
to function.  The “coarse” calibration R12 will allow a
limited range of scaling values.  The meter can be
scaled down to 1/2 the value of the input, but not
scaled up.  Maximum scaling is maximum input or a
maximum reading of 19999, which ever is lower.
Example: A 2 volt input has a maximum reading of
1.9999 counts, so you cant double the 2 volts, but
you can make 1 volt to read 1.9999.    The “fine” cal-
ibration R9 allows for an approximate range of 1% of
the “coarse” calibration.  Apply the full scale input to
the meter.  Adjust R12 to be within 1% of the desired
scaled value, then use R9 to obtain the final desired
result.

Offset Adjustment:

To adjust the offset, apply the off-

set input signal. (e.g. 4 mA on 4-20 mA input).  Use
R7 for a “coarse” adjustment of the offset.  The offset
can be scaled up or down approximately 2500
counts.  Adjust R7 to within 1% of the desired scaled
value, then use R6 to obtain the final desired result.

Using a screwdriver or thumbnail,
spread tab on each side of case to
unlock top half.  Lift rear top half
and slide away from front of meter.

FRONT

OF METER

LOCKING

TAB

TOP HALF

OF CASE

R9

R12

SCALE
ADJUST
COURSE

SCALE
ADJUST
FINE

OFSET
ADJUST
COURSE

OFSET
ADJUST
FINE

R7

R6

Decimal Point:

To select a decimal point, connect the appropriate DP pin (DP1, DP2, DP3, or DP4) to DP COMMON (HOLD). Unused DP inputs
may remain unconnected (open).

Reading Hold:

Connect DP4/HOLD to DP/COMMON (HOLD).  If this feature is not required, the HOLD pin may remain unconnected (open).  
Hold is optional on this model, and when so equipped, replaces DP4.  This feature is available from our modification centers.

Input Signal:

Connect the IN HI and IN LO to the signal to be monitored.

Supply Power:

Connect the power to the L1 and L2 terminals.  For AC powered units, L1 and L2 are not polarized.

For 9-32 VDC powered units, L1 must be positive with respect to L2.

(Input Signal) IN HI
(Input Signal) IN LO

DP3
DP2
DP1

HOLD

L1 (POS) (Power)

L2 (NEG) (Power)

EX (NEG)

EX (POS)

DP COMMON
(HOLD)

These instruments are designed for maximum safety to the operator when mounted in a panel according to

instructions. They are not to be used unmounted or for exploratory measurements in unknown circuits.

Range

Resolution

Voltage

Max Input

Drop

(unfused)

4-20mA

0.1%

200mV

100mA

Range Resolution

Input

Maximum

Impedance

Input

1-5VDC

0.1%          10M

 

750V

0-10VDC

0.1%

10M

750V

0-100VDC

0.1%

10M

750V

INPUTS: DC Process

DP4/

Note: Any physical damage to the meter
during calibration will void the warranty