6
5.
FREQUENCY: 50-133 Hz, no correction required
RESPONSE:
25 Hz, add 2% to reading
6.
METER SIZE:
3 inches
7.
DIMENSIONS:
Height:
5-7/8" (149mm)
Width:
3" (76mm)
Depth:
3-1/2" (64mm)
8.
WEIGHT:
2-1/2 lbs. (1.13Kg)
2.
INSTALLATION
2.1
Unpacking
Examine the shipping carton for signs of damage before unpacking. If shipping
carton is in good condition, unpack and inspect the Instrument for damage in-
curred during shipment. If damaged, notify the carrier and supplier, and do not
use the Instrument. If the Instrument appears to be in good condition, read
Operator’s Manual in its entirety. Become familiar with the Instrument as in-
structed in the manual; then check the electrical performance.
Save the shipping carton and packing materials for future storing or shipping of
the Instrument.
2.2
Power Source Requirements
No supply source is required. The Instrument is energized by the circuit being
measured.
2.3
Installation
The Instrument can be set horizontally or vertically; however, horizontal position
will assure maximum accuracy.
3.
CONTROLS, CONNECTORS & INDICATOR
All operating controls, connectors and indicator are shown in Figure 3-1. Be-
come familiar with each before operating the Instrument for the first time.
3.1
Description
Description of the controls, connectors and indicators are listed in Table 3-1. The
numbers in Figure 3-1 correspond with the numbered items listed in Table 3-1.
Table 3-1. Controls, Connectors and Indicator
1.
15 Amp Jack (red dot):
Load side of current coil for 15 Amp range. Also
common for 1500-3000 watt range wattmeter
connection.
2.
3 Amp Jack (red dot):
Load side of current coil for 3 amp range. Also
common for 300-600 watt range wattmeter con-
nection.