9
5.
DC MICROAMPERES:
Range:
0-50
μ
A
Voltage
Drop:
250mV
6.
DC MILLIAMPERES:
Ranges:
0-1-10-100-500 mA
Voltage Drop (Approx.):
250 mV, 255 mV, 300 mV, 500 mV
7.
DC AMPERES:
Range:
0-10A
Voltage Drop (Approx.):
255mV
NOTE:
The 10A range is
not
internally fused.
8.
9.
*ACCURACY:
DC Voltage Ranges:
2% of Full Scale
DC Current:
0-50
μ
A Range: 1.5% of Full Scale
Other Ranges:
2% of Full Scale
**AC Voltage
Ranges:
3% of Full Scale
Frequency Response:
Referenced to 100 Hz (Figures 4-5 and 4-9)
Resistance Ranges:
R X 1:
2.5° of Arc
R X 100:
2°of Arc
R X 10,000:
2° of Arc
10. DECIBELS:
Range:
-20 to +10 dB, -8 to +22 dB,
0 to +30 dB, +6 to +36 dB,
+20 to +50 dB
Reference Level:
With zero dB power level equal to 1 mW across
a 600 line.
11. BATTERIES:
Voltage:
1.5V, 9V
NEDA No.:
13F, 1604A
12. FUSE:
F1 1A, 250V, type 3AG, quick-acting Littlefuse Type
312001, F2 2A, 600V Littlefuse Type BLS or
2A 600V, Bussman Type BBS.
13. TEST LEADS:
1 red, 1 black, 48" long.
14. SIZE:
5½” x 7" x 3 “ (13.34 x 17.78 x 7.94 cm)
15. WEIGHT:
3 lbs. (1.359 kg)
16. ***RATED-CIRCUIT
TO-GROUND
VOLTAGE:
1000V AC/DC Max.
:
E
C
N
A
T
S
I
S
E
1
R
X
0
R
0
1
X
0
R
0
0
,
0
1
X
R
e
g
n
a
0
R
0
0
,
2
-
0
0
0
0
,
0
0
2
-
0
M
0
2
-
0
g
n
i
d
a
e
R
r
e
t
n
e
2
C
1
0
0
2
1
0
0
0
,
0
2
1
n
e
p
O
l
a
n
i
m
o
N
e
g
a
tl
o
V
ti
u
c
ri
C
V
5
.
1
V
5
.
1
V
9
t
r
o
h
S
l
a
n
i
m
o
N
t
n
e
r
r
u
C
ti
u
c
ri
C
A
m
5
2
A
1
m
5
2
.
1
A
μ
5
7
Содержание 260-8P
Страница 1: ...Simpson 260 Series 8P Volt Ohm Milliammeters INSTRUCTION MANUAL...
Страница 3: ...3 NOTES...
Страница 5: ...5 5 2 Battery Replacement 27 5 3 Fuse Replacement 27 5 4 Test Lead Inspection 28 5 5 Care 28...
Страница 29: ...29 NOTES...
Страница 30: ...30 NOTES...
Страница 31: ...31 NOTES...