Operating manual - English
Optika
8
Copyright Silca 2017
4
DEVICE INSTALLATION AND PREPARATION
The device can be installed by the purchaser and does not require any special skills.
However, some checks and preparation for use have to be carried out by the operator.
4.1
Separate Parts
Power supply cable and power supply unit
Connect OPTIKA to the power supply unit (M) and connect to the power source by means of the power
supply cable (fig. 9).
Fig. 9
4.2
Connection to external supply points
For the protection of the device and the operator’s safety it is extremely important to ensure that it is
connected to the power mains with the right voltage and by means of a properly earthed differential
switch.
4.3
Environmental conditions
The OPTIKA device uses a series of instruments whose operation could be significantly affected by the
environment in which they work. The most suitable conditions for the device are:
•
Temperature b10° C e +40° C
•
Relative humidity 60% (Max) (without condensation)
•
It is in any case advisable to avoid excessively damp or badly ventilated areas.
Note: like any device using lenses for reading images, whenever the temperature goes suddenly
from low (0°C or lower) to high (20-25°C), the lenses may cloud and condensation form. In such
cases the device must be placed in an area with constant temperature of 20°C if possible, until the
condensation on the lenses evaporates and the device can operate in normal conditions.
M
power supply