Troubleshooting
Instruction Manual VisGuard 2
14162E/2
97
FAULT MESSAGE
DESCRIPTION
POSSIBLE CAUSES
LIGHTSOURCE 1
The detector for monitoring
the light source receives no
light from the corresponding
light source.
Defective light source.
Service technician
IO PORT
The connection between the
NG_Haupt and NG_Bedi print
in the SICON has been distur-
bed.
Cable disconnected.
Plug connection defective.
POWERBOX
Actuation of the power box
has been disturbed.
Connection to the power box
has been interrupted.
MASTER SW VERS
This fault message is display-
ed when the software version
of the SICON-C is older than
the version of the connected
photometer.
Software is not the most re-
cent version. In this case, the
software of the control unit
must be updated to the most
recent version.
Reference Manual
Table 4: Possible fault messages