‡ SIGNET 2760 DryLoc™ pH/ORP Preamplifier

SAFETY INSTRUCTIONS
1. Depressurize and vent system prior to installation or removal.
2. Confirm chemical compatibility before use.
3. Do not exceed maximum temperature/pressure specifications.
4. Wear safety goggles or faceshield during installation/service.
5. Do not alter product construction.
6. When using chemicals or solvents care should be taken and appropriate eye,

face, hand, body, and/or respiratory protection should be used.

Technical Data

General

Compatible Electrodes:

+GF+ SIGNET DryLoc
pH and ORP Electrodes

Response Time (includes response time of electrodes):

pH:

<6 s for 95% of change

ORP:

application dependent

Temperature, t:

w/2764, 2774 series:

  20 s

w/2754 :

140 s

w/2756

196 s

Electrical (2760-1, -2 Preamplifiers)

Accuracy:

±2 mV @ 25°C (77°F)

Input Impedance, Z:

>10

11

Supply Voltage:

6 to 10 VDC

Environmental

Enclosure Rating: NEMA 6P/IP68

with electrode and watertight conduit and/or
extension pipe connected

Wetted Materials:

CPVC

Cable:

4.6m/15 ft., 3-cond shielded, 22AWG

Maximum cable:

122 m (400 ft.)

Shipping  Weight:

0.64 kg (1.41 lb)

Operating Temperature:

0°C to 85°C (32°F to 185°F)

Storage Temperature:

-20°C to 85°C (-4°F to 185°F)

Relative Humidity:

0 to 95%, non-condensing
(without electrode connected)

Standards and Approvals

CE

Manufactured under ISO 9001:2000 for Quality and ISO
14001:1996 for Environmental Management

Description
The +GF+ SIGNET 2760 pH/ORP Preamplifiers with DryLoc™ connectors are part of George Fischer Signet's growing arsenal of
sensors for industrial fluid measurement and control.
The 2760-1 and -2 Preampliers connects to three different families of pH and ORP electrodes to provide a stable and relaible signal
input to tradiGF+ SIGNET instruments.
The 2760-3 and -4 models use the same package to link the same electrodes to third-party equipment that read a traditional millivolt
signal for pH measurement without temperature compensation or amplification.

3-2760.090

3-2760.090

(A-3/04) English

Dimensions

Bottom View showing

DryLoc™ connectors

31.75 mm
(1.25 in.)

89.4 mm
(3.52 in.)

70.1 mm
(2.76 in.)

56.6mm
(2.23 in.)

Main Contents

Previous View

pH/ORP Menu

Содержание DryLoc 2760

Страница 1: ...terials CPVC Cable 4 6m 15 ft 3 cond shielded 22AWG Maximum cable 122 m 400 ft Shipping Weight 0 64 kg 1 41 lb Operating Temperature 0 C to 85 C 32 F to 185 F Storage Temperature 20 C to 85 C 4 F to 185 F Relative Humidity 0 to 95 non condensing without electrode connected Standards and Approvals CE Manufactured under ISO 9001 2000 for Quality and ISO 14001 1996 for Environmental Management Descri...

Страница 2: ...the threads to secure the assembly 5 Use 3 4 in rigid or flexible conduit to protect the cable from damage 6 If the cable will be extended beyond the standard 15 ft supplied with the 2760 use a junction box to terminate the wires Note Do not extend the 2760 cable more than 120 m 400 ft In line Assembly and Installation 178 mm 7 in 107 mm 4 2 in 3 4 in Conduit customer supplied Fitting 2764 Differe...

Страница 3: ... 4 in watertight pipe to the top of the 2760 Secure the threaded connection to prevent any leakage For additional defense against possible accumulation of condensation at the back seal area of the 2760 fill the lower 3 4 inches 75 100 mm of conduit or extension pipe with a flexible sealant such as silicone 2760 Submersible Sensor Assembly and Installation 94 mm 3 7 in 107 mm 4 2 in Fill with 3 to ...

Страница 4: ... 2756 DI 3 2757 Standard Flat DryLoc 3 2754 3 2754 HF 3 2755 Connect this terminal to Earth GND to reduce electrical interference White Temp Green Temp Blue Iso Gnd Brown mV input Shield Black V Red V 12 11 10 9 8 7 6 5 18 17 16 15 14 13 12 11 16 15 14 13 12 11 10 9 3 8750 1 3 8750 2 3 8750 3 ProcessPro pH ORP Transmitter Any of these electrodes can be connected to the GF SIGNET ProPoint and Proce...

Страница 5: ...RTD 3 2766 1 3KΩ RTD 3 2764 2 PT1000 RTD 3 2766 2 PT1000 RTD 3 2764 3 300Ω RTD 3 2766 3 300Ω RTD DryLoc Preamplifier 3 2760 1 3 2760 2 White Temperature Green Temperature Blue mV Reference Shield Earth Ground Black V Red V Brown Active mV pH ORP Combination ORP Redox Electrodes 3 2775 10KΩ ID 3 2777 10KΩ ID 3 2775 1 no Temp Comp 3 2777 1 no Temp Comp 3 2755 10KΩ ID 3 2757 10KΩ ID 3 2757 WT 10KΩ ID...

Страница 6: ...6 of 8 2760 DryLoc pH ORP Preamplifier SIGNET Notes Main Contents Previous View pH ORP Menu ...

Страница 7: ...7 of 8 2760 DryLoc pH ORP Preamplifier SIGNET Notes Main Contents Previous View pH ORP Menu ...

Страница 8: ...process connection 3 2764 2 159 000 944 Differential pH Electrode PT1000 RTD 1 in NPT process connection 3 2764 3 159 000 945 Differential pH Electrode 300Ω RTD 1 in NPT process connection 3 2766 1 159 000 949 Differential pH Electrode 3KΩ RTD 1 in NPT process connection 3 2766 2 159 000 950 Differential pH Electrode PT1000 RTD 1 in NPT process connection 3 2766 3 159 000 951 Differential pH Elect...

Отзывы: