‡ SIGNET 2760 DryLoc™ pH/ORP Preamplifier
SAFETY INSTRUCTIONS
1. Depressurize and vent system prior to installation or removal.
2. Confirm chemical compatibility before use.
3. Do not exceed maximum temperature/pressure specifications.
4. Wear safety goggles or faceshield during installation/service.
5. Do not alter product construction.
6. When using chemicals or solvents care should be taken and appropriate eye,
face, hand, body, and/or respiratory protection should be used.
Technical Data
General
Compatible Electrodes:
+GF+ SIGNET DryLoc
pH and ORP Electrodes
Response Time (includes response time of electrodes):
•
pH:
<6 s for 95% of change
•
ORP:
application dependent
•
Temperature, t:
w/2764, 2774 series:
20 s
w/2754 :
140 s
w/2756
196 s
Electrical (2760-1, -2 Preamplifiers)
Accuracy:
±2 mV @ 25°C (77°F)
Input Impedance, Z:
>10
11
Ω
Supply Voltage:
6 to 10 VDC
Environmental
Enclosure Rating: NEMA 6P/IP68
with electrode and watertight conduit and/or
extension pipe connected
Wetted Materials:
CPVC
Cable:
4.6m/15 ft., 3-cond shielded, 22AWG
Maximum cable:
122 m (400 ft.)
Shipping Weight:
0.64 kg (1.41 lb)
Operating Temperature:
0°C to 85°C (32°F to 185°F)
Storage Temperature:
-20°C to 85°C (-4°F to 185°F)
Relative Humidity:
0 to 95%, non-condensing
(without electrode connected)
Standards and Approvals
•
CE
•
Manufactured under ISO 9001:2000 for Quality and ISO
14001:1996 for Environmental Management
Description
The +GF+ SIGNET 2760 pH/ORP Preamplifiers with DryLoc™ connectors are part of George Fischer Signet's growing arsenal of
sensors for industrial fluid measurement and control.
The 2760-1 and -2 Preampliers connects to three different families of pH and ORP electrodes to provide a stable and relaible signal
input to tradiGF+ SIGNET instruments.
The 2760-3 and -4 models use the same package to link the same electrodes to third-party equipment that read a traditional millivolt
signal for pH measurement without temperature compensation or amplification.
3-2760.090
3-2760.090
(A-3/04) English
Dimensions
Bottom View showing
DryLoc™ connectors
31.75 mm
(1.25 in.)
89.4 mm
(3.52 in.)
70.1 mm
(2.76 in.)
56.6mm
(2.23 in.)