8550-1 Flow Transmitter
“- - - - -”
“Pulse Overrun Output1”
“Value must be more than 0”
Open Collector is always
activated
Flow rate exceeds display capability
• Open Collector pulse rate exceeds
maximum of 300 pulses per minute.
• Pulse width set too wide.
K-factors cannot be set to 0.
• Hysteresis value too large
• Defective transmitter
• Increase Flow units time base
• Move flow decimal one place to the right
• Increase Pulse volume setting
• Decrease pulse width setting.
• Reduce system flow rate
Enter K-factor from 0.0001 to 99999
• Change the hysteresis value
• Replace transmitter