background image

2

Signet 3719 pH/ORP Wet-Tap

76 mm

(3.0 in.)

130 mm

(5.13 in.)

49 mm

(1.94 in.)

45 mm

(1.75 in.)

39 mm

(1.52 in.)

20 mm

(0.8 in.)

75 mm

(2.94 in.)

Pipe thread

1

1

/

2

 in. NPT

or

ISO 7-R1

1

/

2

76 mm

(3.0 in.)

39 mm

(1.52 in.)

20 mm

(0.8 in.)

Pipe thread

2 in. NPT

or

ISO 7-R2

34 mm

(1.32 in.)

41 mm

(1.62 in.)

25 mm

(1.00 in.)

Saddle length

Saddle height

Outlet

height

Nominal

Pipe ID

Flange width

 

 

Description 

  Nominal 

Order No. 

(ASTM, NPT) 

   Pipe OD 

Flange Width 

Saddle Length  Saddle Height  Outlet Height

2007-0225 

Saddle, 2

1

/

2

 in. x 1

1

/

2

 in. 

2.875 in. 

4.84 in. 

3.11 in. 

4.29 in. 

0.83 in.

2007-0230 

Saddle, 3 in. x 1

1

/

2

 in. 

3.500 in. 

5.43 in. 

3.43 in. 

4.84 in. 

0.83 in.

2007-0240 

Saddle, 4 in. x 1

1

/

2

 in. 

4.500 in. 

5.98 in. 

3.90 in. 

5.90 in. 

0.91 in.

2007-0260 

Saddle, 6 in. x 2 in. 

6.625 in. 

8.90 in. 

4.49 in. 

8.46 in. 

1.50 in.

2007-0280 

Saddle, 8 in. x 2 in. 

8.625 in. 

11.3 in. 

5.71 in. 

11.3 in. 

1.50 in.

2007-0210 

Saddle, 10 in. x 2 in. 

10.75 in. 

12.5 in. 

7.04 in. 

13.25 in. 

1.25 in.

2007-0212 

Saddle, 12 in. x 2 in. 

12.75 in. 

16.0 in. 

9.68 in. 

14.00 in. 

1.20 in.

.

3.  Dimensions

3719-1 Wet-tap

1

1

/

 in. outlet 

for 2

1

/

 to 4 in. 

saddles

3719-2 Wet-tap

2 in. outlet for 

6 to 12  in. 

saddles

CAUTION!

When using these clamp-on saddle fittings, the system temperature and pressure is limited by the saddle specifications.

The process temperature is limited to 40°C (104°F) maximum when using these saddles.

°C

-10

0

°F

10

20

30

40°C

50

14.5

29

45

58

72.5

87

102

116

psi

2

1

3

5

7

32

14

50

68

4

6

8

bar

122

104°F

68

6 in. PP Saddle

2½, 3, 4, 10 & 12 in. PP Saddles

8 in. PP Saddle

218 mm

(8.57 in.)

2716-WT pH Electrode

2717-WT ORP Electrode

2756-WT(-1) pH Electrode
2756-WTP(-1) plastic pH Electrode
2757-WT ORP Electrode
2757-WTP plastic ORP Electrode

35 mm

(1.4 in.)

30.5 mm

(1.2 in.)

12 mm

(0.47 in.)

Содержание 3719

Страница 1: ...trode 2757 WT DryLoc ORP Electrode 2757 WTP DryLoc plastic ORP electrode Shipping Weight Wet tap assembly 1 2 kg 2 7 lbs Electrode 0 13 kg 0 3 lb Process Connection 3 3719 11 1 in NPT 3 3719 21 2 in NPT 3 3719 12 ISO 7 1 R1 3 3719 22 ISO 7 1 R2 Performance Maximum Flow Velocity 10 ft s Efficiency 97 25 C 77 F Response Time pH 5s for 95 of signal change ORP Application dependent Drift 2 mV per week...

Страница 2: ...in 8 46 in 1 50 in 2007 0280 Saddle 8 in x 2 in 8 625 in 11 3 in 5 71 in 11 3 in 1 50 in 2007 0210 Saddle 10 in x 2 in 10 75 in 12 5 in 7 04 in 13 25 in 1 25 in 2007 0212 Saddle 12 in x 2 in 12 75 in 16 0 in 9 68 in 14 00 in 1 20 in 3 Dimensions 3719 1 Wet tap 11 2 in outlet for 21 2 to 4 in saddles 3719 2 Wet tap 2 in outlet for 6 to 12 in saddles CAUTION When using these clamp on saddle fittings...

Страница 3: ... pH ORP Transmitter 10 20 pH 25 0C ENTER a 3719 pH ORP Wet Tap b Low Profile PP Clamp on Saddle Fitting ASTM sizes 2 to 12 in c DryLoc pH or ORP Electrode 6 versions available see ordering infomation on page 8 DryLoc refers to the electrode connector style d 2750 or 2760 DryLoc pH ORP Preamplifier Sensor electronics e Output signal options S3 L 4 to 20 mA All of these components items a to d are s...

Страница 4: ...of the 3719 system is strongly recommended Choose the 3719 version 11 or 21 appropriate to the size of the branch connection of the required fitting Use 11 for sizes 2 5 to 4 in and 21 for sizes 6 to 12 in 5 3 Installation in Pipe Sizes less than 2 5 Inches It is possible to install the 3719 into pipe sizes below 2 5 inches by creating a flow cell with standard piping components One simple solutio...

Страница 5: ...electrode installed the pipe contents are exposed Electrode Piston 5 5 Electrode Installation 1 Remove the safety plug from top of electrode piston Slide electrode straight down into electrode piston Thread electrode into place until connector shoulder is flush with top of electrode piston Hand tighten only 2 Place the Locking Shroud over electrode turn turn clockwise to unlock the piston then pre...

Страница 6: ...e or sand the glass electrode surface Treat glass electrode surfaces with appropriate care to avoid breakage 6 Electrode Maintenance and Cleaning 5 6 Electrode removal The electrode in any pH or ORP system requires periodic service calibration or replacement When removing the electrode from the wet tap assembly it is very important to exercise caution and follow the instructions carefully 1 Remove...

Страница 7: ... the pH electrode offset The mV offset will track across the entire pH range The slope is usually not affected by offset changes i e pH 7 10 mV pH 4 187 mV slope 59 mV pH Electrode Offset pH 7 buffer 25 C Theoretical pH 7 0 0 0 mV New electrode pH 7 0 25 pH 15 mV Reliable pH 7 0 85 pH 50 mV Electrode offsets greater than 0 85 pH 50 mV indicate the electrode requires cleaning or replacement See Mai...

Страница 8: ... In line Preamplifier with in ISO threads and 4 6 m 15 ft cable 3 2760 31 159 001 369 In line Connector with 4 6 m 15 ft cable and in NPT threads 3 2760 41 159 001 370 In line Connector with 4 6 m 15 ft cable and ISO 7 1R threads 2007 0225 159 000 812 PP Clamp on Saddle 2 5 in x 11 2 in ASTM NPT 2007 0230 159 000 813 PP Clamp on Saddle 3 in x 11 2 in ASTM NPT 2007 0240 159 000 814 PP Clamp on Sadd...