![Signametrics SMX4030 Скачать руководство пользователя страница 17](http://html1.mh-extra.com/html/signametrics/smx4030/smx4030_operators-manual_1268486017.webp)
17
Signametrics
Configuration Option Checks
- This section of the main panel allows the selection of one out of all of the possible
pre defined configurations. On power up, the configuration is set to
Disabled
. Possible selections include
Disabled,
TwoWire, FourWire, SixWire, Universal, TwoGroups, and FourGroups
.
Edit
Menu
- This menu item has three selections, setting all parameters to their default value, opening the Edit
Scanner Parameter sub-panel, and opening the Scan List Editor sub-panels.
Trigger
Menu -
This menu item controls the functionality of the scanner’s trigger input and output signals. It
enables or disabled the Trigger output, and sets the polarity of both Trigger input and output.
Tools
Menu
- This menu item provides access to some of the special test and cleaning tools. It includes the Self-
Cleaning function, the various self-tests, and the shorted input detector.
Operations
Menu -
This menu item allows running some of the special scanner’s operations such as AutoScan,
Triggered Scan, Triggered Auto Scan, and the setting and monitoring of the Trigger output and trigger input
respectively.
4.0 Scanner Tutorial
Most of the SMX4000 functions are accessible from the control panel described in the previous section. This
section describes in detail the Scanner’s operations and practices. To gain familiarity with the SMX4000 series
Scanners, run the Windows ‘SETUP.EXE’ to install the software, then run the Scanner control panel to demonstrate
its operations.
4.1 Scanner Configurations
The various configurations maybe experimented with using the control panel. Connect to the Scanner’s terminals to
the appropriate channels and bus lines, making sure not to exceed any of the specifications and limits. Use the
control panel’s option checks to select the desired operation mode, and the switch buttons to close or open the
appropriate channels.
The SCANSetConfig() DLL function sets the Scanner configuration. The amount of time it takes to set a new
configuration is equat to the set Actuation time. The Scanner can function as a multiplexer, or as an uncommitted
matrix switch. In the
TwoWire
,
FourWire
,
SixWire
,
TwoGroups,
and
FourGroups
configurations the Scanner acts as
a multiplexer. This means that when a channel is selected, all other channels in the group are opened prior to closing
the selected channel. For instance, in
FourWire
configuration, the responose to ‘select Ch2) will result in opening of
a previousely selected channel relays and the closure of Ch2 and Ch18. All this will be done within a single
actuation time.