Model 7225 Firmware Update Pack - Installation Instructions

 

 

SIGNAL RECOVERY

 

SIGNAL RECOVERY

 is part of 

d

f

 Advanced Measurement Technology, Inc 

 

801 SOUTH ILLINOIS AVENUE  

5 ASHVILLE WAY 

OAK RIDGE 

MOLLY MILLARS LANE 

TN 37831-2011 

WOKINGHAM, BERKS RG41 2PL 

USA   

UNITED KINGDOM 

Phone:  

+1 865 482 4411 

Phone:  

+44 (0)1252 556800 

Fax:    

+1 865 481 2410 

Fax:  

+44 (0)1252 556899 

 

E-mail

[email protected]

              Web Site

www.signalrecovery.com

 

09072018

 

 

If using the USB adaptor, plug this into a free USB port, follow the instruction 
for installing the supplied driver from CD, and then connect the other end of the 
adaptor into the RS232 cable.

 

 
c)  Select Start - Run and browse to find the Windows update program 

UPDATE.EXE

, which by default will have been placed in a folder named 

c:\Program Files\Signal Recovery\DSP Update Wizard

, and double-click on 

the   program icon to start it. 

 
d)  When the program starts, simply follow the on-screen instructions. In the event 

that the update is for any reason unsuccessful, turn the instrument off and start 
this procedure again. 

 
e)  The update typically takes approximately 10 minutes although it may take longer 

with slow computers or if the computer is being used for other work while it is 
running. 

 
f)  If the procedure completes successfully then the lock-in amplifier will now be 

updated with the new firmware and you may start using it simply by cycling the 
power to the instrument two times, each time leaving it switched off for 5 
seconds before turning it back on. 

 

Further Advice 

If you require further help or assistance in performing this update then please contact 
our service department via e-mail at 

[email protected]

 