Refrigerated Centrifuge Sigma 1-16K
Refrigerated Centrifuge Sigma 1-16K IVD
10 Technical data
52 / 70
Version 10/2013, Rev. 2.21 of 08/05/2020
• sb
Translation of the original operating manual, part no. 0700302
Manufacturer:
Sigma Laborzentrifugen GmbH
An der Unteren Söse 50
37520 Osterode (Germany)
Type:
1-16K
1-16K IVD
Order no.:
10046, 10047, 10048
10052, 10053, 10054
Connection requirements
Electrical connection:
Protection class:
IP code:
see name plate
I
20
Power consumption (kW):
Input fuse (AT):
0.54 (at 220-240 V / 50/60 Hz)
0.54 (at 120 V / 60 Hz)
0.54 (at 100 V / 50/60 Hz)
10.0 (at 220-240 V / 50/60 Hz)
10.0 (at 120 V / 60 Hz)
10.0 (at 100 V / 50/60 Hz)
Performance data
Max. speed (rpm):
Max. capacity (ml):
Max. gravitational field (x g):
Max. kinetic energy (Nm):
15,000
72
20,627
5,176
Other parameters
Time range:
Temperature range:
10 sec
– 99 min 59 sec, short run, continuous run
-10°C to +40°C
Physical data
Height (mm):
Height with open lid (mm):
Width (mm):
Depth (mm):
Weight (kg):
291
547
310
660
34
Noise level (dB(A)):
54 (at max. speed)
Refrigerant data
Refrigerant:
Filling quantity (kg):
Max. permissible pressure (bar):
CO
2
equivalent (t):
R134a
0.136
23
0.177