Refrigerated Centrifuge Sigma 1-16K
Refrigerated Centrifuge Sigma 1-16K IVD
6 Using the centrifuge
36 / 70
Version 10/2013, Rev. 2.21 of 08/05/2020
• sb
Translation of the original operating manual, part no. 0700302
6.3.10 Rotor selection
In the delivery status of the centrifuge, the rotor 12134 is preselected. If
another rotor is installed the configuration must be changed, so that the
allowed maximum speed of the rotor can be reached.
• To select another rotor, press the set key and hold it for approx.
2 seconds.
• Choose the number of the installed rotor out of the following list, by
pressing the arrow keys:
Code
1
2
3
4
5
Rotor
11024
12024
12120
12134
12135
12136
Code
6
7
8
9
10
Rotor
12137
12180
12118
--
--
• Press the set key to confirm the input.
The RCF value will be adjusted automatically to the selected rotor.
NOTE
The rotor selection will be saved under the corresponding program
number.