Quick Start
SPD3303S/3303D
Programmable DC Power Supply
V1.0
2013 SIGLENT TECHNOLOGIES CO., LTD
Страница 1: ...Quick Start SPD3303S 3303D Programmable DC Power Supply V1 0 2013 SIGLENT TECHNOLOGIES CO LTD...
Страница 2: ...publication replaces all previous corresponding material 3 SIGLENT reserves the right to change the specifications and price of the product 4 Any way of copying extracting and translating the content...
Страница 3: ...ply AC Input Voltages 100V 110V 220V 230V 10 50 60Hz Use proper fuse The fuse types 100V 110V T6 3A 250V 220V 230V T3 15A 250V Make sure to use the correct type of fuse before turning on the instrumen...
Страница 4: ...ge Please keep proper ventilation and check the fan and air vents regularly when using the instrument Operate condition Location indoor no strong light almost no Interfering pollution Comparative humi...
Страница 5: ...icates direct injury or hazard that may happen WARNING Indicates potential injury or hazard that may happen CAUTION Indicates potential damage to the instrument or other property that may happen Symbo...
Страница 6: ...D3303S 3303D 4 3 TFT color LCD display with content of 16M Three independent outputs two of which are adjustable and the total power up to 195W Four kinds of input voltage values includes 100V 110V 22...
Страница 7: ...7 Output check 8 Chapter 2 Control panel operation 9 2 1 Output summary 10 2 2 CH1 CH2 Independent Output 11 2 3 CH3 Independent mode 12 2 4 CH1 CH2 Series mode 13 2 5 CH1 CH2 Parallel mode 15 2 7 Ti...
Страница 8: ...contents of the shipment have been completely checked and the instrument has passed both electrical and mechanical tests The consigner or carrier will be responsible for damages to the instrument res...
Страница 9: ...e DIP Switch at the rear panel according to the actual demand Warning To switch to the wanted power voltage please disconnect the power cord first Electrical Check Please use the power cord provided a...
Страница 10: ...Area 10 Power Switch 3 Model 11 CH1 Output Terminal 4 System parameter configuration button 12 Ground Terminal 5 Multi function knob 13 CH2 Output Terminal 6 Fine Adjust button 14 CV CC indicator lig...
Страница 11: ...setting interface Press the button longer to lock it and shorter to turn to system information interface Buttons for controlling the channel Press the button to turn on off all the channels Press the...
Страница 12: ...on front panel The output terminals of CH1 CH2 and CH3 include positive terminal and negative terminal and a common ground for CH1 and CH2 additional Each channel has its own logo The operation detail...
Страница 13: ...annel logo 3 Operating mode logo The corresponding logo will display when working in CV or CC mode 4 LOCK logo It will display when the LOCK is turned on 5 USB logo It will display when there is a USB...
Страница 14: ...panel Description 1 warning message 2 The DIP switch of the AC power and its identification 3 The description of the AC input voltage 4 AC power socket 5 The fan air vents 6 CE certification mark 7 U...
Страница 15: ...values of all channels are not zero 2 Check the voltage output of CH1 CH2 Turn on CH1 CH2 and the instrument work in CV mode Check whether the voltage value could be changed from 0V to 32V 2 Current o...
Страница 16: ...tion of SPD3303S 3303D control panel will be introduced in detail to giving you an all around understanding of it which will eventually lead to easier work Brief introduction Output summary CH1 CH2 in...
Страница 17: ...e that of the single channel In the series mode the voltage value is twice that of the single channel Constant voltage current In the constant current mode independent or tracking mode the current val...
Страница 18: ...egative terminals of CH1 CH2 3 Set voltage and current value of CH1 CH2 Firstly move the cursor to select the wanted parameter voltage current and then revolve the multi function knob to change the co...
Страница 19: ...3V 5V 3A Operation steps 1 Connect the load to the positive and negative terminals of CH3 on the front panel 2 Select the wanted voltage value by moving CH3 DIP switch 3 Open output Press ON OFF butt...
Страница 20: ...eries mode and the indicator light turns bright immediately 2 Connect the load to the positive terminal of CH2 and the negative terminal of CH1 3 Turn on CH2 and then adjust its current value by revol...
Страница 21: ...SIGLENT 14 SPD3303S 3303D Quick Start Note You can identify the current working state CC or CV by referring to indicator light of CH1 CH2 red means CV green means CC...
Страница 22: ...steps 1 Press the PAR button to start Parallel mode and the indicator light turns bright immediately 2 Connect the load to the positive and negative terminal of CH1 3 Turn on CH1 and then set the outp...
Страница 23: ...SIGLENT 16 SPD3303S 3303D Quick Start...
Страница 24: ...FILE CHOICE by revolving the multi function knob 4 Select OPEN CHOICE by moving the cursor 5 Move the cursor to STORE and then press it to save the current setup Recall setup Operation steps 1 Press...
Страница 25: ...consecutive output and the longest time of each group is 10000s Setup steps 1 Press CH1 CH2 to select the wanted channel 2 Press TIMER to enter the Timer Setup interface and the indicator light turns...
Страница 26: ...n button 2 Revolve the multi function knob to make theTimer state to ON 3 Press the knob to start theTimer Note If press the TIMER button when the timer is running then you can observe the decreasing...
Страница 27: ...button to enter Waveform Display interface and the indicator light turns bright immediately 3 Press CH1 CH2 ON OFF button to turn on CH1 CH2 channel the corresponding indicator light turns bright imm...
Страница 28: ...SIGLENT SPD3303S 3303D Quick Start 21 2 9 Version information Press LOCK button quickly to enter Version Information interface which is shown below...
Страница 29: ...ce 1 Open the EsayPower software after USB line having been connected perfectly and make sure the instrument is connected with the software correctly 2 Click Version and then choose Upgrade in the dro...
Страница 30: ...3D Quick Start 23 Figure 2 Figure 3 4 As it shows in figure 4 click Upgrade button to start upgrading The upgrade is finished when the progress bar complete and the instrument will run the version aft...
Страница 31: ...dure also can be used if the method above is failed Specific steps are below 1 Press the knob before powering on the instrument and the turn on the instrument and it will enter the guide procedure mod...
Страница 32: ...n installed with EasyPower software or NI Measurement Automation 3 1 Command list 1 IDN 2 SAV 3 RCL 4 INSTrument CH1 CH2 5 INSTrument 6 MEASure CURRent 7 MEAsure VOLTage 8 MEASure POWEr 9 SOURce CURRe...
Страница 33: ...are version Example Siglent SPD3303S SPD00001130025 1 01 01 02 2 SAV Command format SAV name Description Save current state in nonvolatile memory with the specified name Example SAV 1 3 RCL Command fo...
Страница 34: ...d channel query the current channel Example MEASure CURRent CH1 Return Info 3 000 Command format MEASure VOLTage CH1 CH2 Description Query voltage value for specified channel if there is no specified...
Страница 35: ...ce CURRent SOURce CH1 CH2 Description Query the current value of the selected channel Example CH1 CURRent Return Info 0 500 Command format SOURce VOLTage value SOURce CH1 CH2 Description Set voltage v...
Страница 36: ...t 1 Series 2 Parallel Description Select operation mode Example OUTPut TRACK 0 Command format OUTPut WAVE SOURce state SOURce CH1 CH2 state ON OFF Description Turn on off the Waveform Display function...
Страница 37: ...nd format TIMEr SOURce state SOURce CH1 CH2 state ON OFF Description Turn on off Timer function of specified channel Instruction The command works effectivly only when secnum starts from 1 Example TIM...
Страница 38: ...nfo into a binary The state correspondence relationship is as follow Example SYSTem STATus Return info 0x0224 Bit NO Corresponding state 0 0 CH1 CV mode 1 CH1 CC mode 1 0 CH2 CV mode 1 CH2 CC mode 2 3...
Страница 39: ...H3 Power coefficient 5mV Load coefficient 15mV Ripple and noise 2mVrms 5Hz 1MHz Tracking Track error 0 5 10mV of Master No Load operation Parallel coefficient Line 0 01 3mV Load 0 01 3mV rating curren...
Страница 40: ...cord 30M or above DC 500V Operating Indoor environment Altitude 2000 m Environment temperature 0 40 Comparative humidity 80 Installation level Pollution level 2 Storage Environment temperature 10 70 e...
Страница 41: ...not 0 Answer 3 Yes because when current output load is higher than the limited current value the working mode turns to CC from CV Question 4 How to deal with the upgrade failed Answer 4 please do it a...
Страница 42: ...S 3303D Quick Start 35 Contact SIGLENT MTR Add 3 F Building 4 Antongda Industrial Zone 3rd Liuxian Road 68 District Baoan District Shenzhen P R CHINA Tel 86 755 36615186 E mail sales siglent com http...