![SIGLENT SDM3065X Скачать руководство пользователя страница 6](http://html.mh-extra.com/html/siglent/sdm3065x/sdm3065x_user-manual_1267453006.webp)
SIG
L
ENT
IV SDM3055 Digital Multimeter
3. Current input (I) terminal
I and LO terminals
are used for current measurement. The maximum
current which go through the
I
terminal
is limited to 10A by the fuse on the
back panel.
NOTE:
Voltage on the current input terminal corresponds to voltage on LO terminal.
To keep good protection, only use the fuse of specified type and level to
replace this fuse.
IEC Measurement Category II Over-voltage Protection
SDM3065X Digital Multimeter provides over-voltage protection for
line-voltage mains connections meeting both of the following conditions to
avoid the danger of electric shock
:
The
HI
and
LO
input terminals are connected to the mains under
Measurement Category II conditions as following.
The maximum line voltage of the mains is 600VAC.
WARNING:
IEC Measurement Category II includes electrical devices connected to mains
at an outlet on a branch circuit, such as most small appliances, test
equipments, and other devices that plug into a branch outlet or socket.
SDM3065X is capable of making measurements with the
HI
and
LO
inputs
connected to mains in such devices (up to 600VAC) or the branch outlet itself.
However, the
HI
and
LO
terminals of SDM3065X
can‟t be connected to
mains in permanently installed electrical devices such as the main
circuit-breaker panels, sub-panel disconnected boxes and permanently wired
motors. Such devices and circuits are prone to exceed the protection limits of
SDM3065X.
NOTE:
Voltages above 600VAC only can be measured in circuits that are isolated
from mains. However, there may be transient over-voltage in circuits that are
isolated from mains. SDM3065X is able to withstand occasional transient
over-voltage up to 4000Vpk.
Please don‟t use this instrument to measure
circuits that transient over-voltage may exceed this level.
Содержание SDM3065X
Страница 1: ...User Manual SDM3065X Digital Multimeter UM06036 E01B 2017 SIGLENT TECHNOLOGIES CO LTD ...
Страница 2: ......
Страница 27: ...SIGLENT SDM3055 Digital Multimeter 15 Temperature Measurement For RTD and thermcouple sensors ...
Страница 78: ...SIGLENT 66 SDM3055 Digital Multimeter Diagram 2 39 System Information ...