SIGLENT cp5050 Скачать руководство пользователя страница 7

 

7/15 

 

 

1

Sensor head 

This clamps the conductor being measured, and carries out the actual current 

measurement. It is a precision assembly including a molded component, a ferrite 

core,and a Hall effect element. It may be damaged if subjected to sudden 

changes in ambient temperature, or mechanical strain of shock, and therefore 

great care should be exercised in handling it. 

2

Opening lever   

Operating lever for opening the sensor head. Always use this lever to open the 

sensor head.   

3

BNC output connector 

Connect to the BNC input connector of the oscilloscope. The output of the 

current probe is terminated internally. You must select the input impedance of 

the oscilloscope to be 1 MΩ in order to make accurate measurements. If the 

oscilloscope you are using does not have a 1 MΩ input impedance setting you 

can purchase 50 Ω to 1 MΩ adapter. 

4

Power indicator LED 

When the power adapter is plugged ,the green LED lights.     

Содержание cp5050

Страница 1: ...Instruction Manual CP5000 Series AC DC Current Probe CP5050 30A DC 40MHz CP5300 150A DC 12MHz Siglent Technologies Co Ltd...

Страница 2: ...are fully understood and met A WARNING notice denotes a hazard It calls attention to an operation procedure practice or the like that if not correctly performed or adhered to could result in personal...

Страница 3: ...Measurement Procedure 8 Attentions 8 Preparations for Measurement 10 Demagnetization and Zero Adjustment 10 Making the Measurement 11 Specifications 11 Service Strategy 14 List of goods 15 Contacting...

Страница 4: ...t terminal BNC is equipped with a protective earthing with double insulation construction Do not allow the device to get wet and do not take measurements with wet hands This may cause an electric shoc...

Страница 5: ...y water or dust proof To avoid damage do not use it in a wet or dusty environment The sensor head is a precision assembly including a molded component a ferrite core and a Hall effect element It may b...

Страница 6: ...up to 30 Amp CP5300 has a 12 MHz bandwidth and is designed to measure continuous currents up to 150 Amp Key features Highly accurate current measurements Easy current measurements Wide bandwidth DC A...

Страница 7: ...in handling it 2 Opening lever Operating lever for opening the sensor head Always use this lever to open the sensor head 3 BNC output connector Connect to the BNC input connector of the oscilloscope T...

Страница 8: ...degaussing and AutoZero about 5 secs 9 Range selected Key When the key is pressed the probe will switch the range The CP5050 has two ranges 30A and 5A The Current Transfer Ratio is 0 1V A and 1V A Th...

Страница 9: ...A 300A Range 5A 1X 30A 10X 30A 10X 150A 100X Over 5A 5A 30A 30A 30A 50A 150A 300A Current Transfer Ratio 5A 1V A 30A 0 1V A 30A 0 1V A 150A 0 01V A Lowest Measurable Current 5A 1mA 30A 5mA 30A 10mA 15...

Страница 10: ...10 15 Fig 1 CP5050 Continuous maximum input rating Frequency derating Fig 2 CP5300 Frequency response Fig 3 CP5300 Continuous maximum input rating Frequency derating...

Страница 11: ...r Diameter Max 5mm 20mm CableLength 1m 1 5m BNC Length 50cm Adapter Dimensions 72 62 31mm Clamp Dimensions L W H 75 40 18mm 175 68 29mm Termination Unit L W H About 119 49 28mm Probe Weight 240g 500g...

Страница 12: ...ture Depending on the measured current frequency some sound maybe produced by resonance but has no effect on measurements The reading may be affected by the position within the clamp aperture of the c...

Страница 13: ...xcess current is input generated heat activates a built in safety function that blocks normal output If this happens remove the input immediately remove the sensor from the conductor being measured or...

Страница 14: ...000A series connects the power supply the red LED lights Range select and Degauss AutoZero 1 With the oscilloscope input at ground adjust the trace to the zero position 2 Set the input coupling of the...

Страница 15: ...mly locked and the sensor head securely closed Service Strategy Problems Possible cause Processing methods Can t measuring DC Power off Power on Oscilloscope set AC mode Oscilloscope set AC mode Senso...