CP6000/CP4000
Current Probe Instructions 25
Specifications
Model
CP4020
CP4050
CP4070
CP4070A
BW
(
-3dB
)
DC-200kHz
DC-1MHz
DC-300kHz
DC-300kHz
Rise time
≤1.75uS
≤0.35uS
≤1.2uS
≤1.2uS
Continuous
maximum
input range
20Arms
50Arms
70Arms
70Arms
Max peak
current value
60A
140A
200A
200A
Range switch
50mV/A
500mV/A
50mV/A
100mV/A
5mV /A
50mV /A
5mV /A
10mV /A
DC Accuracy
(Typical)
±2%
(
0.4A-10ApK
)
at 50mV/A;
±3%±20mA
(
20mA-14ApK
)
at 500mV/A
;
±2%
(
0.4A-10ApK
)
at 50mV/A
±3%±50mA
(
50mA-10ApK
)
at 100mV/A
;
±2%
(
1A-60ApK
)
at 5mV/A
±4%±200mA
(
200mA-100ApK
)
at 50mV/A
;
±2%
(
1A-200ApK
)
at 5mV/A
±4%±50mA
(
500mA-40ApK
)
at 10mV/A
;
±15% max
(
100A-140ApK
)
at 50mV/A
;
±15% max
(
40A-100ApK
)
at 100mV/A
;
Power
Supply
9V battery
Maximum
Working
Voltage
300V CAT III
600V CAT II
Maximum
Floating
Voltage