Appendix E.
Errors and Warnings
The table below lists some of the errors and warnings:
Code
Error/warning text
Comment/tip
E10
Battery level low
Charge the battery
E11
Battery level critical
Charge the battery
E12
Battery level empty
Charge the battery
E14
Battery imbalance
No problem, will balance out.
E16
Battery temperature high
Contact Sigicom Support
E18
Battery is missing
Check that both batteries are
installed
E19
Backup battery low
Contact Sigicom Support
E43
Sensor test error
Check that the instrument is
correctly aligned. If it is, contact
Sigicom Support.
E81
Monitoring started automatically
After instrument reboot
E86
Instrument alignment incorrect
The instrument shall be mounted
horizontally
E90
Real-time clock not synchronized
Check instrument clock
12
E91
Real-time clock poorly
synchronized
Check instrument clock and set if
it differs two seconds or more.
E100/1 Temperature alarms
Cool down the instrument
E102
Instrument humidity warning
Check that all screws of the
instrument lid are tightened
12
Is set if instrument time differs two seconds or more from correct time,
see section 5.8.3