During Insulation resistance measurement the             symbol 
flashes as long as the high voltage is present at the terminals

5. Insulation Resistance Measurement (M

W

)

5.1. Preparation for measurement

·

Select the required "Test Voltage (50V/100V/250V/500V/1000V)"
by changing the "Function Selector Switch".

·

Connect the test probes firmly to the circuit under test.

Note !

Insulation  resistance  can  only  be  measured  at  voltage  free
object.  Hence  the  circuit  under  test  must  be  switched  off, 
de-energized  and  isolated  before  carrying  out  any  insulation
 resistance measurement.
When  measuring  high-ohmic  insulation  resistance  do  not
 touch the measurement cables.

If  an  interference  voltage  >  25V  is  present  within  the  system,   
insulation resistance measurement is disabled. 
The  instrument  defaults  to  the  default  voltmeter  and  displays  the 
value of the interference voltage on the LCD.
Also the instrument indicates this additionally by an acoustic signal.

HIGH-VOLTAGE

Do  not  touch  the  conductive  ends  of  the  test  probes
after  insulation  resistance  measurement  has  been
activated at the instrument.

A  current  with  a  value  of  2  mA  (limited  by  the  instrument) 
may  flow  over  your  body,  and  although  this  is  not  life 
endangering, the electric shock is distinctly perceptible.
If you are taking a measurement at a capacitive device under test, for 
example  a  cable,  it  may  be  charged  with  as  much  as 
1000V, depending upon the selected nominal voltage. Touching the 
device  under  test  after  measurement  may  be  life  endangering  in 
such cases!

5.2

Insulation resistance measurement

The  instrument  provides  flexibility  for  both  manual  and
continuous  (for  a  pre-selectable  time)  insulation  -  resistance
measurement.

5.2.1 Manual Insulation resistance measurement :

In manual Insulation resistance measurement, Press and hold "TEST" 
key until the display has stabilized. Insulation resistance measurement is 
ended when the yellow TEST key is released. Instrument will continue to 
show the last reading for few seconds even after releasing the "TEST 
key".

Note !

The  instrument's  batteries  are  depleted  during  insulation
resistance  measurement.  Only  press  and  hold  the TEST  key
as  long  as  it  is  necessary  to  take  the  reading  in  the  manual
measurement  mode.  Continuous  measurement  as  described
below should be performed, if absolutely necessary.

7

14

11.1 Description of Menu Items :

11.1.1 Setting Backlit OFF/ON :

  (at all ranges)

Basic

                  

        OFF/ON

Display
1. Now scroll through ON/OFF values using the                key until

you come across required value.

2. To save this value press                 key. After saving, this will take

you to basic display.

3. To go back without saving press 

"ESC"

 key. This will take you to 

main menu "        "

4. To go back to basic display from main menu "        "  use 

"ESC" 

key.

11.1.2 Setting the GoNoGo values :

(at all ranges except V     )
Operating Order :

Basic

                   

  noGo

        nnnn

Display
Setting :

nnnn = noGo limit values in M

/

G  (for Insulation ranges)

nnnn = noGo limit values in   (for Ohm and Continuity)

1. Now scroll through different values using the                key until

you come across required value.

2. To save this value press                 key. After saving, this will take

you to basic display.

3. To go back without saving press 

"ESC"

 key. This will take you to 

main menu "noGo"

4. To go back to basic display from main menu "noGo" use 

"ESC"

 key.

11.1.3 Setting the Timer values

(Only at Insulation Resistance measurement ranges)

Operating Order :
Basic

                             noGo

             ti                   mm:ss

display
Setting :   mm= minutes,  ss = seconds

1. Now scroll through different values using the               key until

you come across required value.

2. To save this value press              key. after saving, this will take

you to basic display.

3. To go back without saving press 

"ESC" 

key. This will take you to

main menu "ti"

4. To go back to basic display from main menu "ti" use 

"ESC"

 key.

11.1.4 To see the stored values

(at all ranges except V     )

Operating Order :
Basic

                       noGo

   ti             reCL             nnnn

display

1. You can now see the first stored value. The scale pointer is used

to indicate the number (out of 10) of the displayed stored value.

2. Now you can view the successive stored value using the

key. The scale pointer also increments to indicate the number
(out of 10) of the displayed stored value.

3. To go back press 

"ESC"

 key. This will take you to

main menu "reCL"

4 To go back to basic display from main menu "reCL" use 

"ESC"

 key.

Ω

Ω

Ω

Содержание Zeta 20

Страница 1: ... Central Buildings Woodland Close Old Woods Trading Estate Torquey Devan England TQ27BB Website www sifamtinsley com uk Contact Number 44 O 1803 615139 E mail info tinsley co uk Sifam Tinsley Instrumentation 3105 Creekside Village Drive Suite No 801 Kennesaw Georgia 30144 Contact Number 1 404 736 4903 Web www sifamtinsley com P R E C I S I O N I N S T R U M E N T A T I O N ...

Страница 2: ......

Страница 3: ...mplete warranty card is not presented to authorised person at the time ofrepairs 3 Meter accessories is not used as per the instructions in the instruction manual 4 Defect caused by misuse negligence accidents tampering and Acts of God 5 Improperrepairingbyanypersonnotauthorisedbythecompany 6 Modification Alterationofanysortismadeinelectricalcircuitry 7 Sealprovidedinside outsideisbroken Warranty ...

Страница 4: ...aking capacity of 10kA Recommended fuse manufacturer Siba Part No 70 065 63 0 5A Be absolutely sure that only the specified fuses are used The use of a fuse with different triggering characteristics a different nominal current or a different breaking capacity exposes the operator the system and the measuring instrument to danger The use of repaired fuses or short circuiting of the fuse holder is p...

Страница 5: ...3 Fuses 19 13 4 Housing 19 Ω Mechanical Design Protection Instrument IP50 For terminal socket IP20 according to DIN VDE 0470 Part 1 EN60529 Dimensions W x H x D 84mm x 195 mm x 35 mm Weight 500 g including battery 13 Maintenance Attention Disconnect the instrument from the measuring circuit before opening the instrument to replace the battery or the fuse 13 1 Battery Before initial Start up or aft...

Страница 6: ...connecting test leads Test leads and crocodile clips must be in good condition clean and with no broken or cracked insulation Meaning of symbols on the instrument u 4 Warning concerning a point of danger Attention observe documentation Earth Continuous doubled or reinforced insulation EU Conformity Mark 17 Power Supply Battery 6 x 1 5V cells IEC Lr6 Alkaline Manganese rechargeable cells provided i...

Страница 7: ...sing the ON OFF button 2 By changing the function selector switch from current position to desired function Switch on is acknowledged by a sound signal The instrument is now ready for operation Attention Before opening disconnect the instrument from the measuring circuit and ensure high voltage is not available on leads Automatic turn off Your instrument turn off automatically if any of the keys o...

Страница 8: ...0 00 Manual 0 00 0 00 0 00 0 00 0 00 0 00 0 00 12 Specifications Meas Range Resolu Accuracy Overload function tion of rdg value and digit duration 1 For insulation resistance range 2 For low Ohms continuity Ranges Terminal voltage on open Open circuit voltage circuit DC 5V 1V d c 0 to 30 of rated voltage Lead resistance Short circuit current 2 mA compensation 0 9 99 W Test current on load 1 mAat m...

Страница 9: ... key Note The instrument s batteries are depleted during insulation resistance measurement Only press and hold the TEST key as long as it is necessary to take the reading in the manual measurement mode Continuous measurement as described below should be performed if absolutely necessary 7 14 11 1 Description of Menu Items 11 1 1 Setting Backlit OFF ON at all ranges Basic OFF ON Display 1 Now scrol...

Страница 10: ...t continues for the pre determined time Pre selectable time 10 sec 5 min Refer to the 11 Menu structure section for how to set the value for pre selectable time Once the time is set it is applicable to all the 5 ranges of Insulation measurement n Select the required value of the time for which the test is carried out n BrieflyPresstheTESTkey Theinsulationresistancetestcarries on for the time inter...

Страница 11: ...g MIN MAX function The MIN MAX function is switched off when 1 Key is pressed for approximately for 1 Sec OR 2 Then function selector switch is operated OR 3 When the instrument is turned OFF and ON again 9 Function MIN MAX Measured Value MIN and MAX Meter reaction Display Measured value Digital MIN MAX Sound signal 1 Activate and store 2 Store and display 3 Return to 1 Reset short short short sho...

Страница 12: ...alue Default value is 0W Refer to the 11 Menu structure section for how to set the value for the limit value 7 3 Zero adjustment When measuring very low resistance values you can eliminate resistance of leads and contact resistance by zero adjustment Connect the test leads to instrument and connect the free ends of the leads to one another Wait for reading to stabilize Briefly press the TEST push ...

Отзывы: