This Quick Install Guide is Applicable for the InnovaSonic

®

 210. 

A copy of this Quick Install Guide and the 

210 Instruction Manual

 are included on the product information CD included in your 

shipment. The information is also available for 

download

.

 

 

Keypad Functions 

The InnovaSonic 210i keypad offers 12 dual-function keys for operation of the menu system plus six quick-setup keys. For quick setup 
of your InnovaSonic, simply press any of the keys across the top to get the desired menu parameter instantly. 

 

 

Follow these guidelines when using

 

the InnovaSonic 210i menu system:

 

 

Keys 

 to 

  and  

 are the menu codes to input 

information required for flow measurement. 

 

Use 

 to delete characters to the left. 

 

Use 

  and  

 to return to the last menu or open the next menu. 

These keys acts as "+" and "-" when used to enter numbers. 

 

 is used to select a menu using two digit numbers from the menu. To 

enter a selected menu, press 

 key first then press your selected digit 

numbers. Hit enter to save the selected menu. For example, to input a pipe 

outside diameter, press 

 

 keys, where “11” is the 

window ID to display the parameter for pipe outside diameter. 

 

After all data is selected use 

 to confirm your choices. 

 

Use 

 to enter / exit the SD card storage 

interface. 

 

 , , , ,

 

 

are 

shortcuts to the windows for Flow Rate, Velocity, 
Signal Quality, POS Totalizer, Graph, and Diagnosis. 

 

 

Installation Steps 

1.

 

Plug in the transducers (See Figure 2). Open the hinged top 
cover of the electronics. Figure 2 shows the downstream 
transducer connector, upstream transducer connector, the 
battery recharge port (charge the transmitter or connect to a 
standby power supply), and the 4-20mA output connector.  

 
 

Figure 1. InnovaSonic 210i Keypad with Dual Function Keys 

Figure 2.  Transducer Connections 