Chapter 4 Troubleshooting & Repair 

FastFlo™ 620S  Instruction Manual

 

4-2 

IM-62S

 

 

Problem 

Possible Cause 

Solution 

 

 

 

Velocity measurement is er-
ratic or fluctuating 

Very erratic or non-uniform flow 

Follow installation requirements given in 
Chapter 2 

 

 

 

 

Flow meter installed with less than 
required pipe diameters upstream 
and downstream of the sensor 

Follow installation requirements shown in 
Chapter 2 

 

 

 

 

Insertion sensor probe not mounted 
securely 

Sensor probe must be mounted securely 
without vibration 

 

 

 

 

Sensor component broken 

Return to factory for replacement 

 

 

 

 

Malfunction in system electronics 

Return to factory for evaluation 

 

 

 

 

Ground loop 

Check wiring, see Chapter 2 

 

 

 

 

Moisture present in the gas flow 

Install a water trap or filter upstream of 
the flow sensor 

 

 

 

Velocity measurement 
seems too high or low 

Sensor assembly not aligned 
correctly to flow 

Correct alignment with the flow indicator 
pointing downstream 

 

 

 

 

Sensor probe not inserted to the 
proper depth 

Verify sensing point is located on the 
centerline of the pipe 

 

 

 

 

FastFlo™

 electronics programming is in-

correct 

Check all settings, see Chapter 3 

 

 

 

 

Flow profile influences 

Correct with K-factor adjustment or find 
another mounting location 

 

 

 

 

Flow has exceeded the maximum 
range of the flow meter 

Reduce flow below the maximum range 
shown on the meter’s nameplate or con-
tact the factory for re-calibration advice 

 

 

 

 

Extremely turbulent flow 

Do not place the meter near a ventilator, 
static mixer or valve 

 

 

 

No response to flow from 
sensor assembly 

No power 

Turn on power to the flow meter 

 

Low flow cutoff too high 

Correct low flow cutoff programming 
us

ing Sierra’s Smart Interface software 

 

 

 

 

Flow rate below meter’s minimum 
flow rating 

Contact factory for instructions 

 

 

 

 

Microprocessor locked 

Switch off power, wait 10 seconds, 
restore power 

 

 

 

 

Sensor failure 

Return to factory for evaluation 

 

 

 

 

Printed circuit assembly defective 

Return to factory for evaluation 

Содержание FastFlo 620S

Страница 1: ...FastFlo 620S Instruction Manual Table of Contents IM 62S 0 1 FastFlo 620S Insertion Mass Flow Meter Instruction Manual Part Number IM 62S Revision L November 2022...

Страница 2: ...Code 201316 Phone 8621 5879 8521 Fax 8621 5879 8586 COPYRIGHT SIERRA INSTRUMENTS 2014 No part of this publication may be copied or distributed transmitted transcribed stored in a re trieval system or...

Страница 3: ...nnections 2 4 Input Power Wiring 2 4 Output Signal Wiring 2 5 Alarm Output Wiring 2 7 Remote Sensor Probe Wiring 2 8 Range Selection Wiring 2 9 Chapter 3 Operating Instructions Using the FastFlo Elect...

Страница 4: ...ons 2 4 2 4 Output Signal Wiring 2 5 2 5 Isolated 4 20 mA Current Loop Connections 2 6 2 6 Non isolated 4 20 mA Current Loop Connections 2 6 2 7 Isolated Alarm Output Connections 2 7 2 8 Non isolated...

Страница 5: ...device verify the flow meter is not actively monitoring or reporting to any master control system Adjustments to the elec tronics will cause direct changes to flow control settings Caution All flow m...

Страница 6: ...grammable transmitter is easily configured via RS 232 and Sierra s Smart Interface software or through three push buttons built into the device Depending on the probe length the FastFlo 620S is suitab...

Страница 7: ...essories are not discarded with the packing material Do not return any equipment to the factory without a Return Material Authorization RMA see Chapter 4 Technical Assistance If you encounter a proble...

Страница 8: ...as flow The electrical power required to maintain a constant temperature differential is directly proportional to the gas mass flow rate The meter electronics measure this power and convert it into a...

Страница 9: ...Gas Calibration Optional Select one of two factory calibrated flow ranges using a simple external customer supplied single contact closure User Full Scale Flow Rate Field configure from 50 to 100 of...

Страница 10: ...for a smoother output Enclosure Options Flow meter electronics are available mounted directly to the flow body or remotely mounted up to 100 feet 60 meters away The electronics housing may be used ind...

Страница 11: ...from the cali bration temperature Line pressure should not vary more than 50 psi 3 4 bar around the calibrated pressure 2 The location meets the required minimum number of pipe diam eters upstream an...

Страница 12: ...der to achieve accurate and repeatable performance install the flow meter using the recommended number of straight run pipe diameters upstream and downstream of the sensor Example A Upstream 1 Require...

Страница 13: ...within 5 perpendicular to the pipe centerline 6 When installed cap the fitting Run a static pressure check on the connection If pressure loss or leaks are detected repair the connec tion and re test...

Страница 14: ...maximum as shown below Confirm power configuration be fore applying power See the flow meter nameplate for input power rat ing Figure 2 3 Input Power Connections Smart electronics and wiring connectio...

Страница 15: ...Output Wiring The 4 20 mA current loop output can be self powered by the flow meter s power supply non isolated or externally powered isolated requiring a separate 18 to 30 VDC power supply Rload is...

Страница 16: ...n FastFlo 620S Instruction Manual 2 6 IM 62S Figure 2 5 Isolated 4 20 mA Current Loop Connections Figure 2 6 Non isolated 4 20 mA Current Loop Connections 4 20 out V out AUX PWR OUT Jumper Current R l...

Страница 17: ...age at the flow meter s power supply is an acceptable driver voltage for the load connected Take into account that the current used by your alarm loads have to come from the flow meter s power supply...

Страница 18: ...ke wiring connections from a sensor probe junction box to a remotely mounted enclosure see Figure 2 10 Figure 2 9 Remote Electronics Enclosure to Sensor Connections Figure 2 10 Sensor Junction Box to...

Страница 19: ...Wiring If your meter is equipped with an optional second range calibration con nect a contact switch as shown below When the switch is closed the de vice changes to Range 2 Open the switch to return...

Страница 20: ...d review the current settings To make changes at the password prompt press the UP arrow until 11 is displayed press FUNCTION to con tinue For flow meters without the display remove the enclosure cover...

Страница 21: ...es the revision number of the software in a series of 3 digits followed by the range number the range number continues to flash every 3 seconds thereafter Using the Single Digit LED for Programming FU...

Страница 22: ...n Meter serial number Flow meter model Start Up Screens Password Total Reset To view settings press FUNCTION twice skipping the password Tochange settings press FUNCTION enter the password 11 press FU...

Страница 23: ...enter the high or low alarm set point value in engineering units 3 Press FUNCTION to advance to the next option or after 12 seconds of non activity the meter returns to the Run Mode and the new pa ra...

Страница 24: ...Run Mode and the new K factor is in effect Entering a K factor without the LCD Display A K factor value of 1 000 VDC means the output value is not affected and is the factory default setting You may...

Страница 25: ...NCTION to advance to the next option or after 12 seconds of non activity the meter returns to the Run Mode and the new UFS is in effect Changing the User Full Scale without the LCD Display If the FFS...

Страница 26: ...or DOWN button until the DVM indicates the desired voltage as shown in the following table Volts Indicated on DVM Time Response Seconds Volts Indicated on DVM Time Response Seconds Volts Indicated on...

Страница 27: ...t the DVM to voltage mode and connect between Vout and Vout 2 Press FUNCTION enter the password if so equipped Press FUNCTION until Zero Volts appears on the LCD display or a solid 1 appears on the LE...

Страница 28: ...DOWN button until the DVM indi cates between 3 95 and 4 05 mA Set DVM back to voltage mode when adjustment is complete 3 After 12 seconds of non activity the meter returns to the Run Mode and the new...

Страница 29: ...and the calibration variables did not drift shift or change values Important The best way to perform this Validation is to use the SIP134 soft ware shipped with your meter It will guide you through t...

Страница 30: ...r if you prefer to validate one of the ana log output signals move the multimeter connection to Vout Record the resulting output voltage in Table 3 1 If using a 4 20 mA calibrated meter set the multim...

Страница 31: ...am such as WordPad Scroll down to Alpha and Ro values for both sensors Use these in your manual calculations If you do not have ac cess to the software contact the factory and supply the serial number...

Страница 32: ...lues and the Ro and Alpha values from the Calibration Certificate to calculate the temperature for each sensor as follows T R R Alpha x R T degrees Celsius R measured sensor resistance R resistance at...

Страница 33: ...stream of flow 5 Make sure there are no leaks in the line being measured After verifying the factors above follow the troubleshooting procedures out lined on the next page If you need to return the fl...

Страница 34: ...ect alignment with the flow indicator pointing downstream Sensor probe not inserted to the proper depth Verify sensing point is located on the centerline of the pipe FastFlo electronics programming is...

Страница 35: ...ion of the product IMPORTANT When contacting Technical Support make sure you have included the following information The flow range serial number Sierra order number and model number all marked on the...

Страница 36: ...low these easy steps to return your instrument for factory ser vice 1 To obtain a Return Materials Authorization RMA number from the Si erra Instruments to http www sierrainstruments com rma new php t...

Страница 37: ...FastFlo 620S Instruction Manual AppendixA Specifications IM 62S A 1 AppendixA ProductSpecifications...

Страница 38: ...Appendix A Specifications FastFlo 620SInstruction Manual A 2 IM 62S...

Страница 39: ...FastFlo 620S Instruction Manual AppendixA Specifications IM 62S A 3...

Страница 40: ...vided return or rejection of product is made within a reasonable period but no longer than one 1 year for calibration and non calibration defects from date of delivery To assure warranty service custo...

Отзывы: