Series 760S/780S UHP Instruction Manual

Chapter 2 Installation

IM-76S-78S-UHP-B

2-5

Wiring Connections

For NEMA 4X enclosures, use TB2 for power and signal connec-
tions, TB1 is for sensor connections. (The terminal designations are
labeled inside the enclosure cover.)

Wiring connections
inside cover

To access components:
1.  Loosen 4 screws.
2.  Remove cover.

NEMA 4X enclosure  

terminal blocks

TB1

TB2

1
3
5
7
9

11
13
15

2
4
6
8

10
12
14
16

6
4
2

5
3
1

Figure 2-2. Wiring Access NEMA 4X Enclosures

For Hazardous-Area enclosures, use the terminal blocks located in-
side the smaller end of the flow meter enclosure for all wiring con-
nections. (The terminal designations are labeled inside the enclosure
cover.) Make sure to observe all CE compliance requirements for
AC wiring connections given on the next page.

Hazardous-Area enclosure 

terminal blocks

Set screw

Wiring
connections

To access components:
1.  Use 1/16 inch hex key to loosen set screw.
2.  Turn cover counterclockwise to remove.

1
2
3
4
5
6
7
8
9

10

20
19
18
17
16
15
14
13
12
11

Figure 2-3. Wiring Access Hazardous-Area Enclosures

Warning!

To avoid potential electric

shock, follow National Elec-
tric Code safety practices or

your local code when wiring

this unit to a power source

and to peripheral devices.

Failure to do so could result

in injury or death. All AC

power connections must be

in accordance with pub-

lished CE directives.

Содержание 760S Series

Страница 1: ...HP Instruction Manual Table of Contents IM 76S 78S UHP 0 1 760S 780 UHP Series Insertion Mass Flow Meter Instruction Manual Part Number IM 76S 78S UHP Revision G August 2018 GLOBAL SUPPORT LOCATIONS W...

Страница 2: ...21 5879 8521 Fax 8621 5879 8586 COPYRIGHT SIERRA INSTRUMENTS 2017 No part of this publication may be copied or distributed transmitted transcribed stored in a re trieval system or translated into any...

Страница 3: ...y Orientation 2 4 Wiring Connections 2 5 Input Power Wiring 2 6 Output Signal Wiring 2 8 Alarm Output Wiring 2 11 Remote Sensor Probe Wiring 2 12 Range Selection Wiring 2 14 Chapter 3 Operating Instru...

Страница 4: ...nnections Hazardous Area 2 8 2 10 Load Resistance Versus Input Voltage 2 9 2 11 Isolated 4 20 mA Loop NEMA 4X 2 10 2 12 Non isolated 4 20 mA Loop NEMA 4X 2 10 2 13 Isolated 4 20 mA Loop Hazardous Area...

Страница 5: ...60S 780S UHP Instruction Manual Table of Contents IM 76S 78S UHP 0 5 List of Tables 2 1 Pipe Length Requirements for Installation 2 2 3 1 Electronics Validation Results 3 15 3 2 Sensor Validation Resu...

Страница 6: ...ctronics device verify the flow meter is not actively monitoring or reporting to any master control system Adjustments to the elec tronics will cause direct changes to flow control settings Caution Al...

Страница 7: ...microprocessor based transmitter integrates the functions of flow range adjustment meter validation and diagnos tics in a probe mounted or remotely mounted housing Mass flow rate and totalized flow as...

Страница 8: ...sure any spare parts or accessories are not discarded with the packing material Do not return any equipment to the factory without first contacting Sierra Customer Service Technical Assistance If you...

Страница 9: ...rature When power is applied to the flow meter the transducer electronics heats the velocity sensor to a constant temperature differential above the gas temperature and measures the cooling effect of...

Страница 10: ...Dual Gas Calibration Optional Select one of two factory calibrated flow ranges using a simple ex ternal customer supplied single contact closure User Full Scale Flow Rate Field configure from 50 to 1...

Страница 11: ...to 200 feet away The electronics housing may be used indoors or out including wet environments Display options include a 2 x 12 character LCD display of mass flow rate including totalized mass or a s...

Страница 12: ...enient access with adequate clearance Also verify the meter is located where the gas is clean and dry and the meter is calibrated for the gas to be measured 4 When using a CSA FM or EEx approved flow...

Страница 13: ...pstream of the sensor Upstream Straight Pipe Length Requirements 1 Piping Condition 760S UHP 2 3 8 1 2 inch 760S UHP 3 1 to 6 inch 780S UHP 4 1 to 6 inch Single 90 elbow or T piece 1 D 15 D 1 D Reduct...

Страница 14: ...ittings or flanges in the pipeline Fitting components should be blown clean with filtered gas be fore use Mount in a vertical position For horizontal pipelines having a process gas temperature above 3...

Страница 15: ...splay board 1 Use a 1 16 inch hex key to loosen the set screw securing the larger end of the enclosure Turn cover counterclockwise and remove 2 Remove 4 screws and standoffs from the display Release t...

Страница 16: ...ter enclosure for all wiring con nections The terminal designations are labeled inside the enclosure cover Make sure to observe all CE compliance requirements for AC wiring connections given on the ne...

Страница 17: ...All EEx installations must use an approved EEx fitting at both cable entries into the enclosure If conduit seals are used they must be installed with 18 inches of the enclosure The Hazardous Area encl...

Страница 18: ...to 4 43 to 5 31 in lbs 0 5 to 0 6 Nm All EEx installations must use an approved EEx fitting at both cable entries into the enclosure If conduit seals are used they must be in stalled within 18 inches...

Страница 19: ...C 0 10 VDC optional or a calibrated 4 20 mA output signal This linear output signal represents 0 100 of the flow meter s user full scale DC Output Wiring The 0 5 VDC 0 10 VDC optional signal can drive...

Страница 20: ...nce in the loop including the wiring resis tance To calculate Rmax the maximum Rload for the loop use the maximum loop current 20 mA The voltage drop in the loop due to resistance is 20 mA times Rload...

Страница 21: ...urrent R load Common 1 2 15 16 Figure 2 12 Non Isolated 4 20 mA Current Loop Connections 1 2 3 4 5 6 7 8 9 10 20 19 18 17 16 15 14 13 12 11 4 20 out 4 20 out R load Current Hazardous Area Enclosures 1...

Страница 22: ...at the flow meter s power supply is an acceptable driver voltage for the load connected Take into account that the current used by your alarm loads have to come from the flow meter s power supply In...

Страница 23: ...e sensor probe to a remotely mounted flow meter enclosure use only factory supplied cables The electronics sensors and interconnecting cables supplied by Sierra Instruments are calibrated as a complet...

Страница 24: ...Enclosure to Sensor Connections NEMA4XEnclosures RED GREEN ORANGE WHITE BLACK Remote enclosure Sensor probe junction box Note S ens or wire color may vary see label in cov er Figure 2 20 Sensor Junct...

Страница 25: ...Remote enclosure Note S ens or wire color may vary see label in cov er Figure 2 22 Sensor Junction Box to Remote Enclosure Connections Range Selection Wiring To access range selection connect two wir...

Страница 26: ...gle digit LED flashes the revision number of the software in a series of 3 digits followed by the range number The range number continues to flash every three seconds thereafter Record Factory Set Par...

Страница 27: ...For units with the optional display you may program the meter without opening the enclosure using the magnetic switches to enter the desired system settings NOTE When activating magnetic switches vari...

Страница 28: ...ter model Start Up Screens Password Total Reset To view settings select FUNCTION twice skipping the password To change settings select FUNCTION At the password prompt use the UP arrow until the number...

Страница 29: ...n User Full Scale K Factor Range in use Software version shown in series of 3 digits 1 Run Mode Press FUNCTION to view or change settings Use the UP or DOWN button to enter new parameters Press FUNCTI...

Страница 30: ...N to enter the high or low alarm setpoint value in engineering units 3 Select FUNCTION to advance to the next option or after 12 seconds of non activity the meter returns to the Run Mode and the new p...

Страница 31: ...to the Run Mode and the new K factor is in effect Entering K factor using the Single Digit LED A K factor value of 1 000 VDC means the output value is not af fected and is the factory default setting...

Страница 32: ...FUNCTION to advance to the next option or after 12 seconds of non activity the meter returns to the Run Mode and the new UFS is in effect Changing the User Full Scale using the Single Digit LED If the...

Страница 33: ...using the Single Digit LED 1 Set the DVM to voltage mode and connect between Vout and Vout on the flow meter terminal block Select the desired range Press the FUNCTION button until a solid 9 appears...

Страница 34: ...elect the desired range Select FUNCTION enter the pass word Select FUNCTION again until Total Reset appears on the display 2 Select the UP button and then the DOWN button until the dis play reads Rese...

Страница 35: ...10 VDC to 0 0 VDC 1 Set the DVM to voltage mode and connect between Vout and Vout 2 Select FUNCTION enter the password if so equipped Select FUNCTION again until Zero Volts appears on the LCD display...

Страница 36: ...LED Adjust UP or DOWN until the DVM indicates between 3 95 and 4 05 mA Set DVM back to voltage mode when adjustment is complete 3 After 12 seconds of non activity the meter returns to the Run Mode and...

Страница 37: ...he calibration variables did not drift shift or change values To perform the instrument validation procedures you will need the following equipment certified digital multimeter with minimum 4 characte...

Страница 38: ...VDC of the bridge voltage point 7 Record the resulting flow shown on the optional LCD display in Table 3 1 If not using a display or if you prefer to validate one of the analog output signals move the...

Страница 39: ...w meter Allow a 6 minute cool down before continuing 3 Remove the cover of the flow meter enclosure to access the sensor connection points Remove the four position jumper from J5 J6 J7 and J8 see belo...

Страница 40: ...d resistance values and the Ro and Alpha Ro values from the Calibration Certificate to calculate the tempera ture for each sensor as follows T R R Alpha x R T degrees Celsius R measured sensor resista...

Страница 41: ...are no leaks in the line being measured After verifying the factors above follow the troubleshooting proce dures outlined on the next page If you need to return the flow meter to the factory see page...

Страница 42: ...gh or low Sensor assembly not aligned correctly to flow Correct alignment with the flow indicator pointing downstream in the direction of flow Flow conditioning plates are not upstream of the sensor C...

Страница 43: ...in the data sheet contact Customer Service at 800 866 0200 or 831 373 0200 in the US or 31 0 20 6145810 in Europe Return shipments to USA Headquarters Sierra Instruments Service Department 5 Harris Co...

Страница 44: ...Series 760S 780S UHP Instruction Manual Appendix A Specifications IM 76S 78S UHP A 1 Appendix A Product Specifications...

Страница 45: ...Appendix A Specifications Series 760S 780S UHP Instruction Manual A 2 IM 76S 78S UHP...

Страница 46: ...Series 760S 780S UHP Instruction Manual Appendix A Specifications IM 76S 78S UHP A 3...

Страница 47: ...Appendix A Specifications Series 760S 780S UHP Instruction Manual A 4 IM 76S 78S UHP...

Страница 48: ...ducts online on Sierra s website Online registration of all of your Sierra products is required for our warranty process Register now at www sierrainstruments com register Lifetime Limited Warranty On...

Отзывы: