Chapter 2 Installation
Series 760S/780S UHP Instruction Manual
2-2
IM-76S-78S-UHP-B
Unobstructed Flow Requirements
Select an installation site that will minimize possible distortion in the
flow profile. Valves, elbows, control valves and other piping compo-
nents may cause flow disturbances. Check your specific piping condi-
tion against the examples shown below. To achieve accurate and re-
peatable performance, install the flow meter using the recommended
number of straight run pipe diameters upstream of the sensor.
Upstream Straight Pipe Length Requirements
(1)
Piping Condition
760S UHP
(2)
3/8 & 1/2-inch
760S UHP
(3)
1 to 6-inch
780S UHP
(4)
1 to 6-inch
Single 90° elbow or T-piece
1 D
15 D
1 D
Reduction (4:1)
1 D
15 D
3 D
Expansion
3 D
10D - 45D
3 D
After a control valve
3 D
45 D
3 D
Two 90° elbows
(in same plane)
3 D
30 D
3 D
Two 90° elbows
(different planes)
5 D
50 D
5 D
(1) Number of diameters (D) of straight pipe required between upstream disturbance and the meter.
(2) Requires 1D of straight pipe downstream of the flow meter.
(3) Requires 3D of straight pipe downstream of the flow meter.
(4) Requires 0D of straight pipe downstream of the flow meter.
Table 2-1. Pipe Length Requirements for Installation