5. Remote Electronics 

Sierra Thermal flow transmitters with E3 in the model code have remote electronics. They have 
a cable with Ex glands between the sensor and the electronics. With these models it is not 
possible or allowed to disconnect the cable from the sensor side. It is possible to disconnect the 
cable and re-

mount it using the electronics’ side.  

 
Please observe the following: Ensure that the power to meter has been disconnected before you 
open the EEx d housing. Then open the terminal housing (short dome on the housing) This 
dome is sealed with a set-screw in the edge of the dome. Unscrew the set-screw and screw the 
dome from the housing. Record the color of the wires to terminals 16 thru 20. Disconnect the 
sensor wires (See diagram below). 
 

 

Fig 2 Wiring diagram remote probe (E3 option) 

 
Unscrew the cable gland and remove the wire from the electronics housing. 
 
To re-install the cable, insert the cable in the cable gland. Then connect the wires to terminals 
16 

– 20, ensuring that they are connected identically to what they were before you removed the 

cable. 
 
Assemble the cable gland according 

to the cable gland manufacturers’ instructions (See 

enclosed), close the housing, tighten the set-screw, and power the system up. 
 
NOTE: The unit has some built in electronic wire length compensation. It is possible to shorten 
the cable or to extend it up to +- 25% of the cable length. Use only Sierra cable PN 23-0221. 
 

Содержание 640S-EEX

Страница 1: ...Appendix for ATEX Certified EX Units Instruction Manual Sierra Instruments Model 640S EEX and 780S EEX Flow Transmitters Part Number IM ATEX Rev C 1 October 2013...

Страница 2: ...nsible for cleaning the mass flow meter to the degree required for your oxygen flow application COPYRIGHT SIERRA INSTRUMENTS 2013 No part of this publication may be copied or distributed transmitted t...

Страница 3: ...he ATEX Approved Equipment 6 4 2 Mounting Commissioning and Operation 6 4 3 Explosive Hazardous Area 6 4 4 Special Conditions for a Safe Use 7 4 5 Particular Recommendations Closing the Cover 7 5 Remo...

Страница 4: ...pril 1994 KEMA ATEX official label from Sierra Instruments Inc The following information is provided as part of the labeling of the transmitter Name and address of the manufacturer Sierra Instruments...

Страница 5: ...ea in which an explosive gas atmosphere is likely to occur in normal operation occasionally Zone 2 Area in which an explosive gas atmosphere is not likely to occur in normal operation and if it does o...

Страница 6: ...ed changes may arise For this reason the instrument must be installed connected operated and maintained according to the instructions in this and the specific product operating manual Persons handling...

Страница 7: ...SME B1 20 1 NOTE Please insure that when you mount these cable glands they are made up wrench tight At least 31 2 turns of the thread must be engaged inside of the electronics enclosure 4 4 Special Co...

Страница 8: ...ew in the edge of the dome Unscrew the set screw and screw the dome from the housing Record the color of the wires to terminals 16 thru 20 Disconnect the sensor wires See diagram below Fig 2 Wiring di...

Страница 9: ...e external earthing connection is located on the boss on the outside of the housing and consist of an 316SS screw 10 24 UNC 2B tread seal spring washer and cable shoe crimp lug suited for maximum 3 mm...

Страница 10: ...r Service at www sierrainstruments com or 831 373 0200 in the USA or 31 0 72 5071400 in Europe Return shipments to USA Headquarters Sierra Instruments Inc Attn Customer Service Department 5 Harris Cou...

Страница 11: ...nds are for indoor use and offer IP54 environmental protection A termination suitable for EMC protection can be made using armoured cables with these glands Clamp options allow woven steel wire and st...

Страница 12: ...on For lead sheathed cable the lead sheath must pass through the Continuity Washer when installation is complete B Expose armour approx 20mm long 5 Slide Cone onto inner sheath and under armour Slide...

Страница 13: ...or taper threads use thread sealant Also see page 1 diagram and Hole Data above 5 The surface of the enclosure should be sufficiently flat and rigid to make both the IP joint and earth contact where n...

Страница 14: ...tection code nn year of manufacture Protection Concept and Gas Groups EEx d IIC EEx e II CENELEC ATEX Ex d IIC Ex e II IEC Ex d Flameproof IIC suitable for Gas Group IIC e g hydrogen ignitable gas air...

Страница 15: ...1 These glands are certified with one specific size of flameproof sealing ring per gland size as supplied 2 These glands must not be used with enclosures where the temperature at the point of mounting...

Отзывы: