Start Up
8.7 Load cell replacement
SITRANS WW310
Operating Instructions, 03/2014, A5E33490433-AA
31
8.7.3
Calibration
Step 1
Power up the integrator and using the appropriate integrator menu, verify that the new load
cell(s) are operating correctly.
Step 2
Zero calibrate the unit using the integrator manual for reference. Record zero calibration
procedure results.
Step 3
Span calibrate the unit using the integrator manual for reference. Record span calibration
procedure results.
Step 4
Re-check zero calibration. Run another zero calibration and record the results. Verify that the
zero repeated to the same value recorded in Step 2 (Page 31).
Step 5
Confirm calibration with material test as described in the integrator manual, if possible, before
placing the feeder back in operation.
Содержание SITRANS WW310
Страница 1: ...SITRANS Weighfeeders SITRANS WW310 Operating Instructions 03 2014 ...
Страница 8: ...Introduction 1 1 The Manual SITRANS WW310 6 Operating Instructions 03 2014 A5E33490433 AA ...
Страница 10: ...Safety notes 2 1 Safety and General Precautions SITRANS WW310 8 Operating Instructions 03 2014 A5E33490433 AA ...
Страница 12: ...Technical data 3 1 Specifications SITRANS WW310 10 Operating Instructions 03 2014 A5E33490433 AA ...
Страница 14: ...Operation 4 1 Weighfeeders SITRANS WW310 12 Operating Instructions 03 2014 A5E33490433 AA ...
Страница 19: ...Mechanical installation 5 1 Installation SITRANS WW310 Operating Instructions 03 2014 A5E33490433 AA 17 ...
Страница 20: ......
Страница 22: ...Calibration 6 3 Span SITRANS WW310 20 Operating Instructions 03 2014 A5E33490433 AA ...
Страница 24: ...Wiring SITRANS WW310 22 Operating Instructions 03 2014 A5E33490433 AA ...
Страница 36: ...Troubleshooting 9 1 Mechanical troubleshooting SITRANS WW310 34 Operating Instructions 03 2014 A5E33490433 AA ...
Страница 47: ......