SITRANS TH/TR/TF
8
A5E32520700-01, 06/2013
Securing the transmitter:
in the connection head cover
in the connection head base
①
Transmitter
②
Ceramic base of the measuring element
③
Connection head
Note
Using the lock washers (TH200/TH300 only)
The lock washers
④
included in the delivery are only required for securely fastening the transmitter when the transmitter
②
is directly installed on the round plate
③
for a temperature probe.
①
Fixing screw M4x35
②
Transmitter
③
Round plate
④
Lock washer DIN 6799
- 3.2 A2
3.1.2
SITRANS TH installation on DIN rail and G rail
Note
Fixing rings
The fixing rings included in the scope of delivery for the transmitter are not required for the installation on DIN rails or G rails.
You can either install the transmitter on a 35
mm DIN rail or on a 32
mm G rail. DIN
EN
60715 applies to DIN rails and
G
rails in this context. The DIN rail adapter required for installation can be ordered as an accessory under the Order No.
7NG3092-8KA.
Adhere to the ambient conditions specified in the technical data.
Содержание SITRANS TF Series
Страница 1: ...1 English 3 Български 34 Čeština 66 Suomi 97 Slovenčina 128 Slovenščina 159 ...
Страница 2: ......