Connecting
4.3 Nameplates for hazardous area installations
SITRANS LU180
20
Operating Instructions, 06/2016, A5E37100674-AA
●
Under certain extreme circumstances, the non-metallic parts incorporated in the enclosure of
this equipment may generate an ignition-capable level of electrostatic charge. Therefore the
equipment shall not be installed in a location where the external conditions are conducive to
the build-up of electrostatic charge on such surfaces.
●
In addition, the equipment shall only be cleaned with a damp cloth. This is particularly
important if the equipment is installed in a zone 0 location.
4.3
Nameplates for hazardous area installations
Note
Always refer to nameplates on the device for device-specific approvals.
ATEX certificate on nameplate
The ATEX certificate listed on the nameplate can be downloaded from the LU180 product page
on our website http:wwww.siemens.com/SITRANSLU180.
Go to Support > Approvals > Certificates.
The IECEX certificate listed on the nameplate can be viewed on the IECEx website:
1.
Go to http://iecex.iec.ch/
2.
Click on Certified Equipment
3.
Enter the certificate number IECEx 15.0118X
①
ATEX certificate
Содержание Sitrans LU-180
Страница 1: ...Operating Instructions LU 180 Sitrans LU 180 Edition 06 2016 Ultrasonic Transmitters ...
Страница 7: ...Introduction SITRANS LU180 6 Operating Instructions 06 2016 A5E37100674 AA ...
Страница 9: ...Description SITRANS LU180 8 Operating Instructions 06 2016 A5E37100674 AA ...
Страница 37: ...Troubleshooting SITRANS LU180 36 Operating Instructions 06 2016 A5E37100674 AA ...
Страница 41: ...Technical data 8 7 Approvals SITRANS LU180 40 Operating Instructions 06 2016 A5E37100674 AA ...
Страница 43: ...Appendix A 1 Measurement interval SITRANS LU180 42 Operating Instructions 06 2016 A5E37100674 AA ...