Parameter reference
SITRANS LR250 (PROFIBUS PA)
114
Operating Instructions, 08/2014, A5E32221386-AC
Response Rate (2.3.8.1.)
Note
Changing Response Rate resets Fill Rate (2.3.8.2), Empty rate (2.3.8.3), and Filter Time
Constant (2.6.8.1.).
Sets the reaction speed of the device to measurement changes.
Response Rate (2.3.8.1.)
Fill Rate (2.3.8.2.)
Empty rate (2.3.8.3.) Filter Time Constant
(2.6.8.1.)
*
Slow
0.1 m/min (0.32 ft/min)
10 s
Medium
1.0 m/min (3.28 ft/min)
10 s
Fast
10.0 m/min (32.8 ft/min)
0 s
Use a setting just faster than the maximum filling or emptying rate (whichever is faster).
Fill Rate (2.3.8.2.)
Defines the maximum rate at which the reported sensor value is allowed to decrease. Allows you
to adjust the SITRANS LR250 response to decreases in the actual material level. Fill Rate is
automatically updated whenever Response rate (2.3.8.1.) is altered.
Options
Range: 0 to 99999 m / min.
Response rate (2.3.8.1.)
Fill rate
* Slow
0.1 m/min (0.32 ft/min)
Medium
1.0 m/min (3.28 ft/min)
Fast
10.0 m/min (32.8 ft/min)
Altered by:
Response rate (2.3.8.1.)
Related parameters
Level unit (2.3.2.)
Enter a value slightly greater than the maximum vessel-filling rate, in units per minute.
Empty Rate (2.3.8.3)
Defines the maximum rate at which the reported sensor value is allowed to increase. Adjusts the
SITRANS LR250 response to increases in the actual material level. Empty Rate is automatically
updated whenever Response Rate (2.3.8.1.) is altered.
Options
Range: 0 to 999999 m / min.
Response Rate (2.3.8.1.)
Empty Rate
* Slow
0.1 m/min (0.32 ft/min)
Medium
1.0 m/min (3.28 ft/min)
Fast
10.0 m/min (32.8 ft/min)
Altered by:
Response Rate (2.3.8.1)
Related parameters
Level Unit (2.3.2.)
•
Enter a value slightly greater than the vessel’s maximum emptying rate, in units per
minute.
Linearization (2.4.)
Volume (2.4.1.)
Carries out a volume conversion from a level value.
Содержание Sitrans LR250
Страница 1: ...Operating Instructions SITRANS LR250 PROFIBUS PA 08 2014 Edition Radar Transmitters ...
Страница 42: ...Installing mounting 4 5 Disassembly SITRANS LR250 PROFIBUS PA 40 Operating Instructions 08 2014 A5E32221386 AC ...
Страница 108: ...Remote operation 7 2 Operating via FDT SITRANS LR250 PROFIBUS PA 106 Operating Instructions 08 2014 A5E32221386 AC ...
Страница 166: ...Parameter reference SITRANS LR250 PROFIBUS PA 164 Operating Instructions 08 2014 A5E32221386 AC ...
Страница 172: ...Service and maintenance 9 4 Disposal SITRANS LR250 PROFIBUS PA 170 Operating Instructions 08 2014 A5E32221386 AC ...
Страница 288: ...LCD menu structure SITRANS LR250 PROFIBUS PA 286 Operating Instructions 08 2014 A5E32221386 AC ...
Страница 289: ...LCD menu structure SITRANS LR250 PROFIBUS PA Operating Instructions 08 2014 A5E32221386 AC 287 ...
Страница 290: ...LCD menu structure SITRANS LR250 PROFIBUS PA 288 Operating Instructions 08 2014 A5E32221386 AC ...
Страница 291: ...LCD menu structure SITRANS LR250 PROFIBUS PA Operating Instructions 08 2014 A5E32221386 AC 289 ...
Страница 292: ...LCD menu structure SITRANS LR250 PROFIBUS PA 290 Operating Instructions 08 2014 A5E32221386 AC ...
Страница 304: ...Index SITRANS LR250 PROFIBUS PA 302 Operating Instructions 08 2014 A5E32221386 AC ...
Страница 305: ......