SITRANS LH100 (7MF1572..) Transmitter for hydrostatic level
A5E32588610-02, 02/2014
35
8
Technical specifications
Gauge pressure input
Measured variable
Hydrostatic level
Measuring range, max.
operating pressure (according
to 97/23/EC Pressure
Equipment Directive) and max.
test pressure (according to
DIN 16086)
Measuring range
Maximum operating
pressure MAWP (PS)
Maximum test pressure
0,0 ... 2.5 bar
7.5 bar
7.5 bar
2-wire output
Output signal
4 … 20 mA
Load
Resistor R [
Ω
]
U
H
Auxiliary power supply in V
Measuring accuracy (according to EN 60770-2)
Reference conditions
●
Rising characteristic curve
●
Start-of-scale 0 bar
●
Room temperature 25 °C (77 °F)
Measurement deviation with
limit setting, including
hysteresis and repeatability
0.3% of full-scale value (typical)
Effect of ambient temperature
Zero point and measuring
range
●
< 6 mH
2
O (3 to 18 ftH
2
O) 0.45%/10 K (0.45%/18 °F) of full-scale value
●
≥
6 mH2O (
≥
18 ftH2O) 0.3%/10 K (0.3%/18 °F) of full-scale value
Long-term stability
Zero point and measuring
range
●
< 6 mH
2
0 (3 to 18 ftH
2
0) 0.25% of full-scale value/year
●
≥
6 mH
2
0 (
≥
18 ftH
2
0)
0.2% of full-scale value/year
Effect of auxiliary power supply In percent per change in voltage
0.01% per 1 V
Rated conditions
Installation conditions
Ambient conditions
●
Ambient temperature
–
Altitude
–
Relative humidity
-10 … +80 °C (-4 … +176 °F)
max. 2 000 m mean sea level
Use a suitable power supply at an altitude of more than 2 000 m (6562 ft.) above sea
level.
0 ... 100 %
Note
Observe the temperature class in hazardous areas.
Storage temperature
-40 … +80 °C (-40 … +176 °F)
●
Degree of protection
according to EN 60529
IP68
Содержание SITRANS LH100
Страница 1: ...1 Deutsch 3 English 22 Français 40 Italiano 59 Español 77 Русский 95 ...