SITRANS FS230
A5E38755607-AA, 02/2017
15
4.4
HI Precision mount
Single enclosure Reflect mount wiring
Connect sensor cables to the transmitter as follows:
1. Fill connector end with Super Lube prior to connecting.
2. Apply Super Lube to the internal threads at the large end of the thread connector.
3. Observing the upstream and downstream orientation
③
, connect upstream sensor cable
②
to transmitter port (1A) and
make connection snug.
4. Connect downstream sensor cable
④
to the transmitter port (1B) and make connection snug.
①
Wall mounted transmitter
③
Flow direction
②
Upstream sensor cable to internal DSL port
(1A) Path 1
④
Downstream sensor cable to transmitter port (1B) Path 1
⑤
Hi Precision mounted sensors
Figure 4-3
Hi Precision Reflect mount single enclosure wiring
5. Tighten all cable glands to obtain optimum sealing.