Nameplate layout
SITRANS FPS200
Operating Instructions, 08/2019, A5E46744245-AB
11
4
Nameplate layout
Example
①
Product name
⑥
Manufacture date
②
Article number
⑦
Serial number
③
Tag
⑧
Maximum allowable test pressure
④
Material
⑨
Maximum allowable operating temperature
⑤
Pipe inner diameter
⑩
Maximum allowable operating pressure
5
Installing/connecting
5.1
General information
The following national regulations have to be observed at installation, especially:
•
The regulations of the pressure equipment directive 2014/68/EU (if relevant).
•
Further applicable standards as AD 2000 or DIN EN 13480 (if relevant).
•
EC machinery directive 2006/42/EC (if relevant).
•
Before assembling/disassembling, the tube or the pipe has to be unpressurized.
•
Before assembling/disassembling, the tube has to be cleaned in case of media that may be toxic or
hazardous to health.
The primary element is calculated for specific pipe lines and operating data. Check before installation if the data
from the calculation according to ISO 5167 matches the actual operating data.