SITRANS F M MAGFLO
®
®
®
®
®
38
SFIDK.PS.026.E5.02
ENGLISH
Error
Error text
#Comment
Outputs
Input
No.
Remedy text
status
status
1
I1 -
Power on
OK
Power on has happened
Active
Active
2
I2 -
Add-on module
Applied
A new module has been applied to the system
Active
Active
3
I3 -
Add-on module
Install
An add-on module is defect or has been removed.
This can be an internal add-on module
Active
Active
4
I4 -
Param. corrected
OK
A less vital parameter in the transmitter has been re-
placed by its default value
Active
Active
20
W20 -
Totalizer 1
Reset manually
During initialisation the check of the saved totalizer
value has failed. It is not possible to rely on the
saved totalizer value anymore. The totalizer value
must be reset manually in order to rely on future
readings
Active
Active
20
W20 -
Totalizer 2
Reset manually
During initialisation the check of the saved totalizer
value has failed. It is not possible to rely on the
saved totalizer value anymore. The totalizer value
must be reset manually in order to rely on future
readings
Active
Active
21
W21 -
Pulse overflow
Adj. pulse settings
Actual flow is too big compared with pulse width and
Reduced
volume/pulse
pulse width
Active
22
W22 -
Batch timeout
Check installation
Duration of batching has exceeded a predefined
Batch out-
max. time
put on zero
Active
23
W23 -
Batch overrun
Check installation
Batch volume has exceeded a predefined maximum
Batch out-
overrun volume
put on zero
Active
24
W24 - Batch neg. flow
Check flow direction
Negative flow direction during batch
Active
Active
30
W30 -
Overflow
Adj. Q
max.
Flow is above Q
max.
settings
Max. 120 %
Active
31
W31 -
Empty pipe
Pipe is empty
Zero
Active
40
P40 -
SENSORPROM
®
Insert/change
SENSORPROM
®
unit not installed
Active
Active
41
P41 -
Parameter range
Switch off and on
A parameter is out of range. The parameter could not
be replaced by its default value. The error will dis-
appear at the next power-on
Active
Active
42
P42 -
Current output
Check cables
Current loop is disconnected or the loop resistance
is too big
Active
Active
43
P43 -
Internal error
Switch off and on
Too many errors occured at the same time
Some errors are not detected correctly
Active
Active
44
P44 -
CT SENSORPROM
®
SENSORPROM
®
unit has been used as CT version
Active
Active
60
F60 -
CAN comm. error
Transmitter/AOM
CAN bus communication error. An add-on module, the
display module or the transmitter is defect
Zero
Inactive
61
F61 -
SENSORPROM
®
error
Replace
It is not possible to rely on the data in SENSOR-
PROM
®
unit anymore
Active
Active
62
F62 -
SENSORPROM
®
ID
Replace
The SENSORPROM
®
unit ID does not comply with the
product ID.
The SENSORPROM
®
unit is from another type of
product MASSFLO
®
, SONOFLO
®
etc.
Zero
Inactive
63
F63 -
SENSORPROM
®
Replace
It is not possible to read from the SENSORPROM
®
unit anymore.
Active
Active
70
F70 -
Coil current
Check cables
Coil excitation has failed
Active
Active
71
F71 -
Internal error
Replace transmitter
Internal convertion error in ASIC
Active
Active
5.6.3
List of error numbers
5.
Commissioning
Note
Some error codes might
never appear in the
MAG 6000 I