Siemens SITRANS F C MASSFLO MASS  6000 Ex-d Скачать руководство пользователя страница 3

SITRANS F C MASSFLO

3

SFIDK.PS.028.Q1.02

Mount the sensor on a vibration-free wall or
steel frame as shown.
Ensure that the sensor is not emptied of liquid
(during normal operation) otherwise incorrect
measurement will occur.

With low flow, horizontal mounting is recom-
mended, in this position air bubbles are
easier to remove.

Locate the sensor low in the system in order to
avoid an under-pressure in the sensor seperat-
ing air/gas in the liquid.

If the liquid is volatile or contains solid
particles, vertical mounting is not recom-
mended.

Always locate the flowmeter as far away as
possible from components that generate me-
chanical vibration in the piping.

Cross talk between sensors mounted close to
each other may disturb the measurement. To
avoid cross talk never mount more than one
sensor on each frame and mount flexible hose
connections between the sensors as shown.

2.

Installation

Vertical mounting

Vibration

Cross talk

2.  Installation

To facilitiate zero-point adjustment, a valve with
tight shut-off should always be mounted in con-
nection with the sensor as a proper zero-point
setting is essential for a good accuarcy.

Note

Please read and follow the directions stated in
the instruction DKFD.PI.028.P1.52 regarding
build-in of sensor.
The instruction is supplied with the sensor.

Zero-point adjustment

Horizontal mounting

Category 2 equipment

May be installed in zone 1 and zone 2.

Category 1 equipment

Sensor MASS 2100 may be installed in zone 0,
zone 1 and zone 2.

Catetory 2 equipment

Signal converter MASS 6000 Ex-d may 

only

 be

installed in zone 1 and zone 2.

2.1 Compact

2.2 Remote

Содержание SITRANS F C MASSFLO MASS 6000 Ex-d

Страница 1: ...Manual Order no FDK 521H1178 SFIDK PS 028 Q1 02 A5E00253712 SITRANS F C MASSFLO MASS flowmeters Signal converter type MASS 6000 Ex d sensor type MASS 2100 compact or remote mounted 083R9506 s ...

Страница 2: ...warning sign are read and understood before the system is being installed Installation connection commissioning and service must be carried out by personnel who are qualified and authorized to do so It is very important that the same people have read and understand the instructions and directions provided in this manual and that they follow the instructions and directions before taking the equipme...

Страница 3: ...to each other may disturb the measurement To avoid cross talk never mount more than one sensor on each frame and mount flexible hose connections between the sensors as shown 2 Installation Vertical mounting Vibration Cross talk 2 Installation To facilitiate zero point adjustment a valve with tight shut off should always be mounted in con nection with the sensor as a proper zero point setting is es...

Страница 4: ...r mixtures are not likely to exist However if they do occur then in all probability only seldom or for short periods of time Instrument category The figures in brackets refer to IEC Built according to European norm E Explosion protected electrial equipment Ex Ex protection labelling in square brackets refers to Associated electrical equipment as an example II 2G E Ex ia IIC T6 Type of protection o...

Страница 5: ... ib IIC T6 1 M20 black ATEX approved glands power supply cable 2 Earthing terminal connect to potential equialising coils PEC 3 M20 blue ATEX approved glands output and input cable 4 Multi plug connection to sensor EEx ia 5 Display with push buttons MASS 2100 Intrinsic safety EEx ia IIC T3 T6 1 Sensor multi plug connection EEx ia MASS 6000 Ex d and MASS 2100 compact EEx de ia ib IIC T3 T6 ...

Страница 6: ...on Type 6 Production year Production date XX week Y year Example 013302N14 3 Week 14 Year 2003 7 Build up code 1 MASS 2100 sensor 8 PED Pressure Equipment Directive 97 23 EC PED SEP marking indicating that the sensor is produced in accordance with Sound Engineering Practice CE 0200 EN 13480 category II marking indicating that the sensor conforms to PED category II 9 Ambient temperature range 10 In...

Страница 7: ...separate part of the terminal house room and must be covered with the plastic cover The cover is marked Increased Safety The signal converter must always be connected to ground using the internal terminal or external PE ground Notified body 2 Installation Sensor MASS 2100 DI 1 5 DI 3 DI 6 DI 15 DI 25 and DI 40 0539 II 1 G DEMKO 03 ATEX 135252X EEx ia IIC T3 T6 Temperature classes for ambient tempe...

Страница 8: ...t the converter on top of the sensor interface Please make sure that it is correctly oriented note the little tap After fitting it can be turned 0 360 The converter is fastened with 4 allen screws Allen key M4 Remote Ex d version For mechanical mounting refer to instruction DKFD PI 028 T1 52 supplied with the mount ing kit The terminals for inputs outputs and power supply can be accessed by removi...

Страница 9: ...t cables must be secured by i e cable strap so that they remain in position even if the cable terminal screws get loose 7 Power supply terminals EEx e maximal core cable 1 5 mm2 Tighten the screws with 0 5 0 6 Nm Mount the plastic cover marked Increased safety again 8 Input and output terminals EEx ia It is an absolute requirement that the wires terminals of the intrinsically safe circuits cannot ...

Страница 10: ...nd and 1 M20 EEX i gland The mains cable is fed through the black PG gland black indicates increased safety e located in the left hand side as viewed from the front The outputs are fed through the blue PG gland the colour blue indicates intrinsically safe circuit i located on the right hand side According to the Ex document issued use of other glands is permitted provided that these are as a minim...

Страница 11: ...6 termi nal screws connecting the SENSORPROM unit 3 Electrical connection 3 2 Setting of active or passive current out put mode View C EEx d The current output in MASS 6000 can operate in either active or passive mode to make elec trical connection as easy as possible The current output in MASS 6000 is default set to active mode If a passive mode is required a jumper on the converter PCB must be p...

Страница 12: ...he converter insert as shown The label text on the add on module should face upwards as shown The add on module is fitted with the connector facing outwards of the enclosure Only HART or Profibus are alloved to be used Please look into technical data page 11 The add on module can be pressed into posi tion Connect the flat cable connector to the module and power up The add on module is auto matical...

Страница 13: ...FIDK PS 028 Q1 02 3 5 Installation examples Current output in passive mode Frequency pulse output in passive mode HART output 3 Electrical connection Intrinsically safe data ia please look into the Technical data in chapter 4 ...

Страница 14: ... Class 150 1 2 1 2 1 11 2 ANSI B16 5 Class 600 Class 300 1 2 1 2 1 11 2 Dairy screwed connector PN 25 40 3 DIN 11851 DN 10 DN 15 DN 32 DN 40 ISO 2853 BS 4825 part 4 SS3351 25 mm 25 mm 38 mm 51 mm Clamp PN 16 3 ISO 2852 BS 4825 part 3 SMS3016 25 mm 25 mm 38 mm 51 mm Thread ISO 228 1 PN 100 G 1 4 G 1 4 ANSI ASME B1 20 1 PN 100 1 4 NPT 1 4 NPT Cable connection Multiple plug connection to sensor 5 2 0...

Страница 15: ...ls 77 78 11 30 V d c Ri 13 6 KΩ Functionallity Start hold continue batch zero point adjust reset totalizer 1 2 force output freeze output Output characteristics Ui 30 V Ii 3 45 mA Pi 0 10 W Ci 0 nf Li 0 mH Galvanic isolation All inputs and outputs are galvanically isolated isolation voltage 500 volts Cut off Low flow 0 9 9 of maximum flow Empty pipe Detection of empty sensor Density 0 2 9 g cm3 To...

Страница 16: ... the output specification please look at the specification page 11 and the note 1 Important When the terminal 31 32 is in active mode may it only be connected to product there is passive The manual must be followed very carefully Note 1 Active output shall be evaluated as barrier output and the Lo and Co is only valid in that case When the active output for current outputs and puls frequency outpu...

Страница 17: ...tional mode characteristics 0 20 mA 4 20 mA Frequency Pulse output Relay Power supply off Power supply on Error relay No error Error Limit switch or 1 set point 2 set points direction switch Low flow Intermediate flow Reverse flow High flow High flow Forward flow Low flow Batch on digital output 4 Technical data ...

Страница 18: ...ettings and error messsages The upper line is for primary flow readings and will always show either mass flowrate volume flowrate density temperature totalizer 1 or totalizer 2 The line is divided into 3 fields S Sign field P Primary field for numerical value U Unit field The centre line is the title line T with individual information according to the selected operator or setup menu The lowest lin...

Страница 19: ...onds The setup menu will operate in two modes View mode Setup mode View mode is a read only mode The pre selected settings can only be scanned Setup mode is a read and write mode The pre selected settings can be scanned and changed Access to the setup mode is protected with a password The factory set password is 1000 Access to a submenu in the set up menu is gained by the lock key A short press on...

Страница 20: ...SITRANS F C MASSFLO 20 SFIDK PS 028 Q1 02 5 3 Menu overview 5 Commissioning ...

Страница 21: ...d open the respective window ensure that the cursor is positioned below the decimal point Use the SELECT KEY move the comma to the required position Use the CHANGE KEY Units are changed by means of the CHANGE KEY with the cursor placed below the unit selected Select units moving the cursor by means of the SELECT KEY Totalizer 2 is not visible when batch is selected as digital output ...

Страница 22: ...SITRANS F C MASSFLO 22 SFIDK PS 028 Q1 02 5 5 Outputs setting menu Current output Digital output Pulse Digital output Frequency 5 Commissioning The current output must be set off when not used ...

Страница 23: ...vel Also possible through digital output Acceptance level is set in the basic settings menu Error number Also possible through digital output Limit switch direction switch Also possible through digital output Direction flow Select 1 setpoint at zero flow Hysteresis at 5 5 Commissioning ...

Страница 24: ...d very accurately a longer integration time can be selected to obtain a better 0 point measurement Setting the 0 point is carried out by activating Zero Adjust A number of individual 0 point adjustments are then made The 0 point found is shown as Zero Adjust The value Zero Sigma shows the standard deviation of the individual measurements made The standard deviation Zero Sigma must be within a wind...

Страница 25: ... values are outside the limit Check flow zero because there is not zero flow in the sensor Check zeroflow conditions valves pumps etc Active Active 35 W35 Current Out 1 Current output exceeds 120 Ensure that the sensor is Check max settings correctly sized and check max flow setting Active Active 36 W36 Freq Pulse Out1 Freq Pulse output exceeds 120 Ensure that the sensor Check max settings is corr...

Страница 26: ...nstant flow signal 2 Air bubbles in medium 2 Ensure medium does not contain air bubbles 3 Vibrations 3 Ensure that the sensor is mounted on a rigid frame without vibrations 4 Pump noise 4 Ensure that pump frequency is different from resonance frequency of sensor Measuring error Undefined Faulty zero point Make new zero point adjustment P40 No SENSORPROM unit Install SENSORPROM unit F61 Deficient S...

Страница 27: ...5158 25 mm ISO 2853 SS 3351 BS 4825 part 4 screwed connection PN 16 6123 6133 38 mm ISO 2853 SS 3351 BS 4825 part 4 screwed connection PN 16 6146 51 mm ISO 2853 SS 3351 BS 4825 part 4 screwed connection PN 16 6158 Type of contact faces For pipe thread Dairy connection State A only A End Flange Facings DIN 2526 form C PN 40 form E PN 100 ANSI B 16 5 ISO 7005 1 type 11 B End Flange Facings DIN 2512 ...

Страница 28: ... c converter F Configuration Standard 0 Density 1 Brix Plato 2 Fraction flow specified by customer contact Siemens Flow Instruments Z Flow calibration Standard calibration included in sensor 3 flow x 2 points A Standard calibration matched pair 3 flow x 2 points FDK 085F7351 B Customer specified matched pair 5 flow x 2 points FDK 085F7372 C Accredited calibration EN 45001 certificate DANAK matched...

Страница 29: ...DK 083H0216 EEx d remote 1 frq pulse output EEx de ia ib T3 T6 version 1 relay output Wall mounting kit for remote Ex d inclusive sensor cable of 5 m FDK 083H0231 Wall mounting kit for remote Ex d inclusive sensor cable of 10 m FDK 083H0232 Wall mounting kit for remote Ex d inclusive sensor cable of 25 m FDK 083H0233 7 4 Spare parts for MASS 6000 Ex d and MASS 2100 Description Version Supply volta...

Страница 30: ...SITRANS F C MASSFLO 30 SFIDK PS 028 Q1 02 8 Certificates 8 1 EC declaration of conformity 8 0 Download certificates Please go to http www siemens com flow ...

Страница 31: ...SITRANS F C MASSFLO 31 SFIDK PS 028 Q1 02 8 Certificates 8 2 EC type examination certificate ...

Страница 32: ...rovement are always welcomed Technical data subject to change without prior notice The reproduction transmission or use of this document or its contents is not permitted without express written authority Offenders will be liable for damages All rights including rights created by patent grant or registration of a utility model or design are reserved Copyright Siemens AG 04 2003 All Rights Reserved ...

Отзывы: