![Siemens SITRANS DS III series Скачать руководство пользователя страница 22](http://html.mh-extra.com/html/siemens/sitrans-ds-iii-series/sitrans-ds-iii-series_compact-operating-instructions_384652022.webp)
SITRANS P, DS III series (7MF4.33.. 7MF4.34.. 7MF4.35..)
20
A5E03434631-02, 12/2012
Note
Electromagnetic compatibility (EMC)
You can use this device in industrial environments, households and small businesses.
For metal housings there is an increased electromagnetic compatibility compared to high-frequency radiation. This
protection can be increased by grounding the housing, see Chapter "Connecting the device (Page 21)".
Note
Improvement of interference immunity
● Lay signal cables separate from cables with voltages > 60 V.
● Use cables with twisted wires.
● Keep device and cables in distance to strong electromagnetic fields.
● Use shielded cables to guarantee the full specification according to HART.
● Refer to HART communication information in Chapter "Technical specifications (Page 36)".