Sensor technical data
Rated conditions
Permissible operating temperature
-40 °C ... 110 °C ( -40 °F ... 230 °F)
Degree of protection according to EN 60529
IP66/IP68
Mechanical load
Class 4M7 to DIN EN 60721-3-4
Climate class
Class 4K4 to DIN EN 60721-3-4
Construction
•
Piezoceramic sensor with pre-amplifier
•
Encapsulated electronics
•
4-wire cable with anti-kink sleeve
Housing material
Stainless steel 1.4571 (316Ti SST)
Cable
Ends with connector sleeves and cable lug
for connection to SITRANS DA400
Weight
125 g (0.276 lb)
Dimensions (W x H x D) in mm (inches)
26 x 29 x 40 (1.02 x 1.14 x 1.57)
Certificates and approvals
Protection against explosion
Intrinsic safety "i"
TÜV 05 ATEX 2876 X
Identification
Ex II 1 G EEx ia IIC T6/T5/T4 or
Ex II 1 D Ex iaD 20, 21, 22 T160
Permissible ambient temperature
Temperature class
Temperature range
T6
-20 °C to +50 °C
T5
-20 °C to +60 °C
•
Use as category 1 G device
T4
-20 °C to +60 °C
T6
-40 °C to +65 °C
T5
-40 °C to +80 °C
•
Use as category 2 G device
T4
-40 °C to +110 °C
•
Use as category 1 D device or as
category 2 D device
T160
-40 °C to +110 °C