Page 38
D00073611 Siemens SD 270 Service Manual
Guidance and manufacturer’s declaration – electromagnetic immunity (1)
The SD 270 Audiometer is intended for use in the electromagnetic environment
specified below. The customer or user of the SD 270 Audiometer should
assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment –
guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be
wood, concrete or
ceramic tile. If floors
are covered with
synthetic material,
the relative humidity
should be at least
30%
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for
power supply
lines
±1 kV for
input/output
lines
±2 kV for
power supply
lines
±1 kV for
input/output
lines
Mains power quality
should be that of a
typical commercial
or hospital
environment
Surge
IEC 61000-4-5
±1 kV
differential
mode
±2 kV
common
mode
±1 kV
differential
mode
±2 kV
common
mode
Mains power quality
should be that of a
typical commercial
or hospital
environment