SM010-3 Siemens SD 270 Operating & Service Manual Page 29
e) If required, readjustment of the input signal level can be accessed by
temporarily deselecting the MASK key - then proceed as in 3.1d)
4.2
Procedure:
As described in 3.2 but adjusting the masking level as required.
IMPORTANT NOTES – FREE FIELD MODES
For the following Free Field modes of operation it is essential for the Free
Field calibration procedure described in Appendix 2 of this operating
manual to have been performed. This aspect may also be subject to local
requirements or legislation.
5. Free Field Live Voice Speech Audiometry
5.1
Set Up:
a) Connect an external amplifier/speaker to the LINE OUT jack socket;
refer to Section 2.7 of this operating manual
b) Connect a microphone to the MIC1 input on the audiometer
c) Press SPEECH & FREEFIELD and use Softkey 1 to ensure that ‘MIC’
is displayed in capitals (indicating that the external microphone is
selected)
d) The external microphone is now routed to the external amplifier and
speaker; use LEFT and RIGHT to select the required amplifier
channel
e) Continue from 1.1d) above
5.2
Procedure:
As described in 1.2.
6. Free Field Recorded Speech Audiometry
6.1
Set Up:
a) Connect an external amplifier/speaker to the LINE OUT jack socket,
and a CD, tape player, or other sound source to the LINE IN jack
socket; refer to Section 2.7 of this operating manual
b) Press SPEECH & FREE FIELD and use Softkey 1 to ensure that
‘LINE’ is displayed in capitals (indicating that the line input is selected)