Siemens AG
SPR2-240.831.01.02.02
POWERMOBIL
07.05
CS PS 24
Maintenance, Operating Value / Functional Inspection
41
Page 41 of 44
Medical Solutions
•
Completely close the slot diaphragm and rotate the slot diaphragm from its basic posi-
tion.
•
Mark the distance and angle of rotation of the displayed slot diaphragm position on the
monitor.
X
•
Release fluoroscopy briefly.
The actual position (distance of the plates and angle of rotation) of the slot diaphragm is
visible.
•
The previously displayed monitor display must match the position and angle of rotation
of the actual slot diaphragm position.
•
Open the slot diaphragm completely and rotate it again.
•
Mark the distance and angle of rotation of the displayed slot diaphragm position on the
monitor.
X
•
Release fluoroscopy briefly.
The actual position (distance of the plates and angle of rotation) of the slot diaphragm is
visible.
The previously displayed monitor display must match the position and angle of rotation
of the actual slot diaphragm position.
NOTE
Perform this check with the lowest possible kV values so that the
front edges of the slot diaphragm plate are effectively visualized.