7ML19985QJ83
Pointek CLS100 – QUICK START MANUAL
Page EN-7
Englis
h
Connections
Cable Version
Enclosure Version
Alarm Output Status
Note:
The mA current loop can be wired in either polarity to determine high or low level
operation as shown in the examples beginning on page 9.
The cable shield should be connected to terminal 3 on the terminal block along with the
provided ground wire.
Alarm Status
Covered
Yellow LED ON
Uncovered
Yellow LED OFF
Power
Connection
high
(fail-safe)
red LED OFF
4 mA
SSS
1
= open
1.
Solid State Switch (SSS)
red LED ON
20 mA
SSS = closed
black wire + V
high
(non fail-safe)
red LED ON
20 mA
SSS = closed
red LED OFF
4 mA
SSS = open
red wire + V
low
(fail-safe)
red LED ON
20 mA
SSS = closed
red LED OFF
4 mA
SSS = open
red wire + V
low
(non fail-safe)
red LED OFF
4 mA
SSS = open
red LED ON
20 mA
SSS = closed
black wire + V
white
red
black
white
Solid state switch, normally
open unpowered.
V supply / mA 2-wire current
loop detection 4/20 or
20/4 mA
Symbols
– ground
Terminal Operation
mA current loop (+V or -V)
mA current loop (-V or +V)
Cable Equivalent
red wire
black wire
cable shield
*relay available only on Pointek CLS100 fully synthetic enclosure
version
solid state switch/relay*
solid state switch/relay*
white wire
white wire
– ground
cable shield connection
Содержание pointek CLS 100
Страница 1: ...Capacitance Switches POINTEK CLS100 Compact Operating Instructions 02 2012 ...
Страница 2: ......
Страница 54: ... EL 4 Pointek CLS100 7ML19985QJ83 Pointek CLS100 4 4 ...
Страница 56: ... EL 6 Pointek CLS100 7ML19985QJ83 9 Pointek CLS100 4 I j 4 100 mm H 6 50 mm 2 50 mm 2 50 mm 2 50 mm 2 ...
Страница 61: ...7ML19985QJ83 Pointek CLS100 EL 11 4 4 6 DC 6 6 óåëßäá 7 4 DC CLS100 4 4 pnp CLS100 4 4 npn ...
Страница 62: ... EL 12 Pointek CLS100 7ML19985QJ83 B CLS100 LED 6 LED 6 4 6 4 ON OFF 4 ON 6 mA 4 óåëßäá 7 ON CLS100 20 ...