4.6.5 Code 15 : JR < JRS
START
CONFIRM mAS LINK ON
D800 PCB
SWITCH OFF THE UNIT
CHECK 'I' FOR 0.6V
+/ -0.1V AC RMS
Check for 22+/- 3 Ohms between
X8.5 and X8.6 and replace D800 if
found faulty.
KV build up OK?
No
Check Inverter ckt/
Inverter pulses/
Resonance cap. C5.
Yes
Voltage betw. X8.5
and X8.6 is OK?
No
Check and
replace D800 if
found fauly.
Check and replace
D915 if found faulty
Yes
STOP