Service Manual
Service Instructions
Siemens Ltd. Med Division India
Version 4.0
Page 24 of 48
Tighten the supplementary lock turning the knob.
Detach the collimator from single tank. Also remove the allied connector connected
to the single tank bottom cover.
Remove the handle, single tank bottom cover and the single tank top cover.
Disconnect the Single Tank connections from Control by removing snap-on
connections and the M5 hex nuts of U, V and Gnd.
Remove the 4 nos. socket head screws M6 X 18 using 5mm Allen key. Loosen the
two socket head screws M10 X 35 using the 8mm Allen key.
Lift the single tank off vertically and place it on a soft base.
Install the new single tank in the reverse order.
Check alignment of light field to radiation field and adjust if necessary.
The parameter adjustments and the calibration of the Unit is Specific to the Single
Tank. In case if the Single Tank is replaced
it is necessary to recalibrate the
equipment.
3.18 Replacing the collimator
In case of any damage to collimator housing, the collimator has to be replaced
completely.
Proceed as described in the following
Loosen the allied connector on the collimator cable and disconnect it from the single
tank bottom cover.
Remove the collimator after loosening the four screws.
Attach the new collimator, fasten it with the four screws Connect the allied back.
Follow the light field to radiation field alignment procedure as mentioned above.
3.19 Replacing the Collimator lamp
Loosen the screw of the clamp, which has marked for light indication on the
collimator. Take out the defective lamp and replace it by a new lamp.
Caution
: Do not touch the glass envelope with your bare fingers.
Check functioning of the lamp.