MAMMOMAT 3000 Modular
Register 5
SPB7-230.092.11
Page 2 of 4
Siemens-Elema AB
Service
Rev. 01
04.98
SPS-UD
Solna, Sweden
2 - 2
Disassembly
2.
Lift the holder (4) and carefully push it inside the collimator.
Removal of diaphragm-plate assembly
2
1.
Loosen the diaphragm-plate assembly (5) by unscrewing four Allen screws (6).
2.
To release the diaphragm-plate assembly, cut the cable ties on each side of the
tube head
.
3.
Lower the diaphragm-plate assembly and let it hang in the cables (7).
5
6
MAM00151
6
Fig 2.3
7
MAM00152
Fig. 2.4