1 - 1
Siemens-Elema AB
Register 5
SPB7-230.092.11
Page 1 of 2
MAMMOMAT 3000 Modular
Solna, Sweden
Rev. 01
04.98
SPS-UD
Service
Prerequisites
1
Purpose
1
The purpose of this modification is to upgrade a MAMMOMAT 3000 Modular with Mo
X-ray tube to Mo/W X-ray tube.
Tools and expendables required
1
•
Standard assembly tools
•
Circlip pliers
•
Denatured ethanol, 99.5 % pure (may only be denatured with toluene, which does not
leave any residues when evaporated) for cleaning the filters or tube window.
•
Lint-free cloth or Q-tips (cotton tops) for cleaning
•
Lint-free cotton gloves
•
Service PC with connecting cable (99 00 440 RE999)
Parts required
1
Filter-assembly kit, part no. 64 82 967 X041E, comprising the following parts:
Documents required
1
Installation and Start-Up Instructions for MAMMOMAT 3000 Modular.
Time required
1
Approximately four hours for one person.
Important notes
1
•
The Allen screws can be very hard to loosen. When unscrewing, be sure of enough grip
with the Allen screwdriver, so that the screws do not get damaged.
•
The filters in the filter assembly must not be contaminated in any way. Therefore, always
use lint-free cotton gloves during the handling of the filter assembly. Do not remove the
filter assembly from its packing until installation.
•
If the filters have been contaminated, clean by using the specified alcohol. Other types
of solvents may leave residues.
Qty
Part
1
Filter assembly
1
Retaining ring
5
Allen screw M5 (countersunk)
4
Cable tie
1
Adapter cable