3
SITRANS F M MAGFLO
Electromagnetic flowmeter type MAG 5000 CT & MAG 6000 CT
Sealing
MAG 6000 CT is installed as shown in the instructions for a standard MAG 6000, except for the
external sealing.
The external sealing is carried out as shown.
Internal totalizers
Depending on type of approval it is possible to reset the internal totalizers. The type of approval
is selected in the reset menu, with the hardware key mounted. It is possible to select between:
•
Hot/cold water
•
Other media than water
Resetting of totalizers by electrical input are not possible.
Hot/cold water
•
Totalizer 1 is allocated to forward flow (cannot be reset).
•
Totalizer 2 is allocated to reverse flow (cannot be reset).
Other media than water
Both totalizer 1 and totalizer 2 are pre-selected to measure the net forward flow, i.e. any reverse
flow will make the totalizers count backwards.
•
Totalizer 1 can be reset if the flow rate in the meter pipe is <0.25 m/s. When the totalizer is
reset, any pulse output will also be reset.
•
Totalizer 2 is consecutive and cannot be reset.