Section 6: Minor Replacements and Adjustments, Page 6—9
5. Remove each tube from the pump assembly.
a. Push the thumb tab on the pump arm to the left as far as possible.
b. While holding the thumb tab in this position, pull the plastic cuff on the tube until
the tube is completely removed from the pump arm.
c. Discard the tube in an appropriate waste container.
Figure 6-10. Removing Tubing from Pump Assembly
6. Remove the new tubes from the Hematek Pump Tube Set.
Each tube is numerically coded to correspond with the numbers shown on the Stain
Pak.
7. Attach the coded end of the new tube to its respective cannula.
a. Replace the cannula if it appears damaged or bent using the Hematek Cannula Set
(PN 4483).
b. If you are re-using any cannulas, make sure the cannula previously used for stain
is used
only
with the stain tubing.
Tube
Code
stain tube
1
buffer tube
2
rinse tube
3
Содержание Hematek
Страница 1: ......
Страница 6: ...Contents ...
Страница 8: ...Page 1 2 Section 1 Introduction ...
Страница 22: ...Page 2 2 Section 2 Installation ...
Страница 32: ...Page 3 2 Section 3 Operating Instructions ...
Страница 38: ...Page 3 8 Section 3 Operating Instructions ...
Страница 40: ...Page 4 2 Section 4 Specimens ...
Страница 44: ...Page 4 6 Section 4 Specimens ...
Страница 46: ...Page 5 2 Section 5 Maintenance ...
Страница 54: ...Page 5 10 Section 5 Maintenance ...
Страница 56: ...Page 6 2 Section 6 Minor Replacements and Adjustments ...
Страница 76: ...Page 7 2 Section 7 Troubleshooting ...
Страница 82: ...Page 8 2 Section 8 Service Supplies and Replacement Parts ...
Страница 88: ...Page 8 8 Section 8 Service Supplies and Replacement Parts ...
Страница 94: ...Page B 2 Appendix B Hematek Preservice Checklist ...
Страница 98: ...Index 4 ...