Section 6: Minor Replacements and Adjustments, Page 6—3
Section 6: Minor Replacements and Adjustments
This section is provided as an aid for performing minor replacements and adjustments on
the Hematek Slide Stainer. Fully review and understand the procedures before attempting
them, and they must be performed with care. For any adjustments or replacements other
than those given in this section, or if any procedure appears to be too complex, refer to
Supplies and Replacement Parts
on page 8-6, for instructions on service for your
instrument.
Fuse Replacement
Use the following procedure to replace the instrument line fuse for all voltage instruments.
The fuse is 1.0 amp, 5mm x 20mm, Type T (Part No. 40151108).
1. Turn the instrument off and remove the line cord from the AC electrical outlet.
2. Using a small, blade-type screwdriver, pry the fuse cover open.
.
Figure 6-1. Opening the Fuse Cover
Содержание Hematek
Страница 1: ......
Страница 6: ...Contents ...
Страница 8: ...Page 1 2 Section 1 Introduction ...
Страница 22: ...Page 2 2 Section 2 Installation ...
Страница 32: ...Page 3 2 Section 3 Operating Instructions ...
Страница 38: ...Page 3 8 Section 3 Operating Instructions ...
Страница 40: ...Page 4 2 Section 4 Specimens ...
Страница 44: ...Page 4 6 Section 4 Specimens ...
Страница 46: ...Page 5 2 Section 5 Maintenance ...
Страница 54: ...Page 5 10 Section 5 Maintenance ...
Страница 56: ...Page 6 2 Section 6 Minor Replacements and Adjustments ...
Страница 76: ...Page 7 2 Section 7 Troubleshooting ...
Страница 82: ...Page 8 2 Section 8 Service Supplies and Replacement Parts ...
Страница 88: ...Page 8 8 Section 8 Service Supplies and Replacement Parts ...
Страница 94: ...Page B 2 Appendix B Hematek Preservice Checklist ...
Страница 98: ...Index 4 ...