Appendix
A.5 Setup Procedure for Wet-Flow Calibration
FUS1010 IP65 NEMA 4X & IP66 NEMA 7
Operating Instructions, 01/2013, A5E02951520-AC
217
The flow calibration report issued with each wet-flow calibration, includes a flow calibration
"Diagnostic Data Sheet." This data sheet lists the site name and other necessary information
(such as sensor serial number and spacing information), for setting-up the flow meter. A wet-
flow calibration applies to a specific flow meter and set of sensors; identified by serial
numbers on the diagnostic data sheet.
Note
In order for the flow calibration to be valid, the flow computer and sensors being used must
have the same serial numbers as those listed for the site on the Calibration Diagnostic Data
Sheet.
Sensor Installation
1.
Refer to the diagnostic data sheet to find the mounting mode (Direct or Reflect) used
during the wet-flow calibration. Review the sensor installation guidelines in your
Operating Instructions manual.
2.
Refer to the diagnostic data sheet for the sensor spacing index utilized during the flow
calibration. Using the mounting configuration employed during the flow calibration, install
the sensors on the pipe at the above noted spacing positions in accordance with the
instructions provided on the sensor installation drawings.
3.
Attach the sensor cables noting that the cable marked UP (upstream) attaches to the
sensor closest to the source of flow.
Note
Before proceeding further, ensure that the pipe is full of liquid. It is not important at this
point that it be flowing.
Flow meter Setup
Note
The following instructions require the use of the keypad and the menu. The installer should
become familiar with their use before proceeding further.
1.
Switch the flow meter on. Press the <MENU> key.
2.
On multi-channel flow meters, use the arrow keys to select [2 Channel Flow] or [Dual
Path Flow] depending on the mode utilized for the wet-flow calibration.
3.
Use the arrow keys to select either [Clamp-on] or [Clamp-on Spool].
4.
Select [Full Site Setup] and use the <Right Arrow> to select [Channel Setup]; then select
[Recall Site Setup].
5.
Use the <Down Arrow> to scroll to the site name indicated on the Calibration Diagnostic
Data Sheet and then press <ENTER>.
Содержание FUS1010
Страница 83: ...Connecting 5 7 Sensor Wiring FUS1010 IP65 NEMA 4X IP66 NEMA 7 82 Operating Instructions 01 2013 A5E02951520 AC ...
Страница 153: ...Functions 7 14 Reflexor FUS1010 IP65 NEMA 4X IP66 NEMA 7 152 Operating Instructions 01 2013 A5E02951520 AC ...
Страница 224: ......
Страница 225: ......
Страница 226: ......
Страница 227: ......
Страница 228: ......
Страница 229: ......
Страница 230: ......
Страница 231: ......
Страница 232: ......
Страница 233: ......
Страница 234: ......
Страница 235: ......
Страница 236: ......
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ......
Страница 241: ......
Страница 242: ......
Страница 243: ......
Страница 244: ......
Страница 245: ......
Страница 246: ......
Страница 247: ......
Страница 248: ......
Страница 249: ......
Страница 250: ......
Страница 251: ......
Страница 252: ......
Страница 253: ......
Страница 254: ......
Страница 255: ......
Страница 256: ......
Страница 257: ......
Страница 258: ......
Страница 259: ......
Страница 260: ......
Страница 261: ......
Страница 262: ......
Страница 263: ......
Страница 264: ......
Страница 265: ...21614 C 1011NFPS 7 ...
Страница 266: ...1011NFPS 7 C 21614 ...
Страница 267: ...1011NFPS 7 C 21614 ...
Страница 268: ...21614 C 1011NFPS 7 ...
Страница 269: ......
Страница 270: ......
Страница 271: ......
Страница 272: ......
Страница 273: ......
Страница 274: ...21614 C ...
Страница 275: ......
Страница 276: ......
Страница 277: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Страница 278: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Страница 279: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Страница 280: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Страница 281: ......
Страница 282: ......
Страница 283: ......
Страница 284: ......
Страница 285: ......
Страница 293: ...Glossary FUS1010 IP65 NEMA 4X IP66 NEMA 7 228 Operating Instructions 01 2013 A5E02951520 AC ...
Страница 297: ...Index FUS1010 IP65 NEMA 4X IP66 NEMA 7 232 Operating Instructions 01 2013 A5E02951520 AC ...