Chapter 3 — Maintenance and Troubleshooting
15
Scale Factor Verification
1
Compensate for mechanical
backlash and adjust the spindle
micrometer for electrical zero.
2
Adjust gap to electrical zero by
moving the probe.
3
Compensate for mechanical
backlash in the micrometer and
adjust to the start of the linear
range.
4
ISF
Vdc Vdc
0.25
n (mV / m)
n - 1
n
µ
=
−
ASF
Vdc
Vdc
2
(mV / m)
250 m
2250 m
µ
µ
µ
=
−
ISF
Vdc Vdc
0.01
n (mV / mil)
n - 1
n
=
−
ASF
Vdc
Vdc
0.08
(mV / mil)
10 mil
90 mil
=
−
Adjust
Micrometer
to...
Record
Voltages
Calculate Scale Factor
n
µ
m
n
or mil
n
Vdc
n
ISF
n
(Incremental
Scale Factor)
ASF
(Average
Scale Factor)
1
250
10
________
2
500
20
________
________
3
750
30
________
________
4
1000
40
________
________
5
1250
50
________
________
6
1500
60
________
________
7
1750
70
________
________
8
2000
80
________
________
9
2250
90
________
________
________
Multimeter
-3.00
±
0.1 Vdc
460
µ
m
or
18 mil
500
µ
m
or
20 mil
500
µ
m
or
20 mil
200
µ
m
or
8 mil
250
µ
m
or
10 mil
Increments:
250
µ
m
or
10 mil
Multimeter